English
Language : 

C11697MB_15 Datasheet, PDF (1/9 Pages) Hamamatsu Corporation – Mini-spectrometer
Mini-spectrometer
TM series
C11697MB
Trigger function and a high sensitivity
CMOS image sensor included
The C11697MB is a polychromator integrated with optical elements, an image sensor and a driver circuit. Light to be mea-
sured is guided into the entrance port of TM series through an optical fiber and the spectrum measured with the built-in image
sensor is output from the USB port to a PC for data acquisition. The C11697MB is a mini-spectrometer with a high-sensitivity
CMOS image sensor mounted on the optical system platform of the previous type C10083MD. The trigger function accepts
even short-time integration so emission spectra from pulsed light can be measured. The readout time is drastically reduced,
making the C11697MB ideal for LED test on industrial productions lines.
The C11697MB comes supplied with free evaluation software that allows setting measurement conditions, acquiring and
saving data, and displaying graphs. Original measurement software can be designed on an end-user’s side as DLL’s function
specification is disclosed.
Features
Applications
Internal high sensitivity CMOS image sensor
(rivaling that of a CCD)
Trigger function included
High-speed readout (2 ms approx.)
Simultaneous charge integration type
Wavelength conversion factor*1 is recorded in internal memory.
Quality check on LED test line
Measurement of pulsed light emission
*1: A conversion factor for converting the image sensor pixel number into a wavelength is recorded in the module. A calculation factor
for converting the A/D converted count into the input light intensity is not provided.
Optical characteristics
Parameter
Value
Unit
Spectral response range
320 to 1000
nm
Spectral resolution
(FWHM)*2
Wavelength reproducibility*3
8 max.
nm
-0.2 to +0.2
nm
Wavelength temperature dependence
Spectral stray light*2 *4
-0.04 to +0.04
-33 max.
nm/°C
dB
*2: Depends on the slit opening. Values were measured with the slit listed in the table “ Structure”.
*3: Measured under constant light input conditions
*4: When monochromatic light of 650 nm is input, spectral stray light is defined as the ratio of the count measured at the input
wavelength, to the count measured in a region of the input wavelength ±40 nm.
Electrical characteristics
Parameter
Specification
Unit
A/D conversion
16
bits
Integration time
30 to 100000
μs
Interface
USB 2.0
-
Consumption current of USB bus power
250
mA
www.hamamatsu.com