English
Language : 

C11482GA Datasheet, PDF (1/8 Pages) Hamamatsu Corporation – Mini-spectrometer
Mini-spectrometer
TG series
C11482GA
For near IR, integrating optical system,
image sensor and circuit
Hamamatsu TG series mini-spectrometer C11482GA is a polychromator integrated with optical elements and an image sensor.
Light to be measured is guided into the entrance port of the C11482GA through an optical fiber and the spectrum measured
with the built-in image sensor is output from the USB port to a PC for data acquisition. The C11482GA is a palmtop-size unit
and operates on USB bus power. It comes supplied with evaluation software that allows setting measurement conditions,
acquiring and saving data, and displaying graphs. Original measurement software can be designed on an end-user’s side as
DLL’s function specification is disclosed.
Features
High throughput due to transmission grating made of quartz
Highly accurate optical characteristics
No external power supply required
(uses USB bus power)
Compact design for easy assembly
Wavelength conversion factor*1 is recorded in internal memory
Applications
Water content measurement
Optical communication component testing
Film thickness measurement
*1: A conversion factor for converting the image sensor pixel number into a wavelength is recorded in the module. A calculation factor
for converting the A/D converted count into the input light level is not provided.
Optical characteristics
Parameter
Specification
Unit
Spectral response range
Spectral resolution (FWHM)*2
Wavelength reproducibility*3
900 to 1700
nm
7 max.
nm
-0.2 to +0.2
nm
Wavelength temperature dependence
Spectral stray light*2 *4
-0.04 to +0.04
-33 max.
nm/°C
dB
*2: Depends on the slit opening. Values were measured with the slit listed in the table “ Structure”.
*3: Measured under constant light input conditions
*4: When monochromatic light of 1300 nm is input, spectral stray light is defined as the ratio of the count measured at the input
wavelength, to the count measured in a region of the input wavelength ±40 nm.
Electrical characteristics
Parameter
Specification
Unit
A/D conversion
16
bit
Integration time
6 μs to 10 s
-
Interface
USB 2.0
-
Current consumption of USB bus power
350 max.
mA
www.hamamatsu.com