English
Language : 

EDS3-124C Datasheet, PDF (4/12 Pages) Fuji Electric – IN-SITU ZIRCONIA OXYGEN ANALYZER
ZSB
CODE SYMBOLS
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
ZSB
2-
4th digit 1
2
5th digit A
B
6th digit Y
A
B
C
D
E
F
G
H
J
K
L
M
7th digit 1
2
3
4
9th, 10th,
11th digits
0Y0
5A3
5A5
5A7
5A1
5B3
5B5
5B7
5B1
6 D3
6 D5
6 D7
6 D1
ZZ Z
12th digit B
E
13th digit J
E
C
14th digit 1
2
15th digit Y
1
Description
Converter mounting (4th digit)
Wall-mounting type (with indication window)
Self-stand type (with indication window)
Sensor, Power supply, Cal. Gas inlet (5th digit)
For general use without protection cover, 100/115 V AC, 50/60 Hz and inlet for φ6 mm tube
For general use with protection cover, 100/115 V AC, 50/60 Hz and inlet for φ6 mm tube
* Protection cover should be selected when ambient temperature is -10°C or lower.
Sensor cable + calibration gas tube (6th digit)
None
6 m cable (without conduit) + 6 m Teflon tube (φ6/φ4)
10 m cable (without conduit) + 10 m Teflon tube (φ6/φ4)
15 m cable (without conduit) + 15 m Teflon tube (φ6/φ4)
20 m cable (without conduit) + 20 m Teflon tube (φ6/φ4)
6 m cable (with conduit) + 6 m Teflon tube (φ6/φ4)
10 m cable (with conduit) + 10 m Teflon tube (φ6/φ4)
15 m cable (with conduit) + 15 m Teflon tube (φ6/φ4)
20 m cable (with conduit) + 20 m Teflon tube (φ6/φ4)
6 m cable (4-core cable only. Heater cable should be prepared by customer.) + 6 m Teflon tube (φ6/φ4)
10 m cable (4-core cable only. Heater cable should be prepared by customer.) + 10 m Teflon tube (φ6/φ4)
15 m cable (4-core cable only. Heater cable should be prepared by customer.) + 15 m Teflon tube (φ6/φ4)
20 m cable (4-core cable only. Heater cable should be prepared by customer.) + 20 m Teflon tube (φ6/φ4)
* When exclusive-special conduit and heater cable are not used, the following conduit should be used.
Conduit with outer diameter φ23 and inner diameter φ16 as specified in JIS C8411
For preparing two heater cables, their rating should be 3 A or more.
Calibration gas unit (7th digit)
Automatic calibration unit + zero gas + filter regulator
Automatic calibration unit + zero gas + air standard gas
Manual calibration unit + zero gas + filter regulator
Manual calibration unit + zero gas + air standard gas
Flow guide tube (9th, 10th and 11th digit)
(Flange)
(Application)
None
None
SUS304 For general use
SUS304 For general use
SUS304 For general use
SUS304 For general use
SUS316 For corrosive
SUS316 For corrosive
SUS316 For corrosive
SUS316 For corrosive
SUS316 Without high particulate cover
SUS316 Without high particulate cover
SUS316 Without high particulate cover
SUS316 Without high particulate cover
Others
Others
(Length)
None
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
300mm
500mm
750mm
1000mm
Others
(Note)
With blow-back unit
With blow-back unit
With blow-back unit
With blow-back unit
Output signal (12th digit)
4 to 20mA DC
0 to 1V DC
Language (13th digit)
Japanese
English
Chinese
Communication functions (14th digit)
RS232
RS485
Reference gas (15th digit)
None
φ6mm tube joint
4