English
Language : 

EDS5-87 Datasheet, PDF (2/8 Pages) Fuji Electric – OPTICAL PRESSURE TRANSMITTER
FBL···3
Process temperature, Allowable pressure limit:
For details, refer to Fig. 1.
Fill-fluid
Code
Process
temperature
Allowable
pressure limit
Silicon oil
Fluorolube oil
Y, G, N
W, A, D
-40 to +100˚C
-20 to +80˚C
2.7kPa abs
Atmospheric
pressure
Silicon oil
R
-15 to +100˚C 2.7kPa abs
Self-diagnosis: Displayed on indication unit (option) and
transmitted to master station.
Item
Measuring range abnormal
Detecting unit failure
Amplifier abnormal
Battery voltage
Battery voltage low alarm
Host system Indication unit
Remote control function:
See Table 1.
Output signal: Optical digital signal
Power supply: Built-in lithium battery (expected life about
4 years)
Optical cable: Code set type, silica fiber ... core/clad di-
ameter 100/140 µm
Optical connector:
FC connector
Transmission distance:
1.5 km max. (when transmission loss of
optical cable is 4 dB/km)
Damping:
Variable from 0.2 to 32 sec (time constant)
Zero elevation and suppression:
Possible within a range from -0.1 MPa
to maximum span.
Explosion-proof: Intrinsic safety type, JIS ib IIC T3
Ambient temperature:
–30 to +70°C
–10 to +60°C for intrinsic safety explo-
sion-proof type
–20 to +70°C when provided with indica-
tor
–10 to +60°C when filled with fluorolube
oil
Storage temperature:
–40 to +80°C
Performance specifications
(Note)
Accuracy rating
±0.1% when measuring span is 1/10 or more
of maximum span
±(0.05 + 0.005 x maximum span/measuring
span)% when measuring span is less than
1/10 of maximum span
Note: Percent value with respect to measuring span (including
linearity, hysteresis and repeatability in standard 23°C status)
Ambient temperature effect:
Zero shift: ± (0.1 + 0.05 × URL/x )%/ 28°C
Overall shift: ± (0.15 + 0.05 × URL/x )%/
28°C
Where; URL: Maximum span
x : Measuring span
3 times as large as above when the 7th
digit (material) is other than V.
Maximum allowable pressure effect:
Zero shift at maximum span
± 0.2% to maximum allowable pressure
Inclination effect:
0.1 kPa/10° or, when 13th digit code (pro-
cessing, fill fluid) is W, D or A, double the
above, at maximum span
Measurement period:
0.2 sec
Response time: Time constant 0.2 sec (at 23°C)
Dead time; About 0.2 sec
Physical specifications
Material:
Material
Detecting unit
code
Process cover Seal diaphragm Other wetted parts
V SCS14
SUS316L
SUS316
J SCS14
SUS316L · Gold-platedSUS316
H SCS14
Hastelloy-C Hastelloy-C
M SCS14
Monel
Monel
T SCS14
Tantalum
Tantalum Except for
B Hastelloy-C lining Hastelloy-C Hastelloy-CFBL 05
L Monel lining Monel
Monel
U Tantalum lining Tantalum
Tantalum
Environmental protection:
Meets JIS C0920, immersion-proof
(equivalent to IEC IP67 or NEMA 6/6P)
Process connection:
Rc1/4 or 1/4-18NPT (whichever selected
by code symbol)
Oval flange threads 7/16-20UNF
Optical cable connection:
G1/2 or 1/2 -14NPT (whichever selected
by code symbol)
Mounting method:
Mounted on the 50A (2B) pipe with U-
bolt
Finish:
Epoxy-polyurethane double coat,
Color: silver (blue for amplifier case cover)
External dimensions:
See OUTLINE DIAGRAM.
Mass:
4.5 to 5 kg
Orientation of transmission unit:
Indicator unit turnable 90° upward/down-
ward relative to detection unit.
2