EDSX3-137D Datasheet, PDF (1/16 Pages) Fuji Electric – IN-SITU ZIRCONIA OXYGEN ANALYZER
IN-SITU ZIRCONIA OXYGEN ANALYZER
DATA SHEET
ZFK8, ZKM, ZTA
This oxygen analyzer is used to continuously measure
oxygen concentration in combustion exhaust gas of industrial
boilers or furnaces, and is ideally suited for combustion man-
agement and control.
The analyzer system is comprised of the detector and con-
verter coupled together as a complete system. Detector set-
ting configuration includes the detector flow guide tube and
detector sensor. The flow guide tube is inserted directly into
the gas and directs gas to the sensor for measurement. The
converter (ZKM) is comprised of the signal processor, input/
output and communications, display and system controls.
The converter is equipped with advanced functionality
such as performing the sensor diagnostics and sensor recov-
ery function, so the detector can be used within long term
stability.
General-use detector
(ZFK8)
High-temperature detector
(ZTA)
FEATURES
1. Gas sampling device is unnecessary
For quick response, insert the detector directly into the
flue Gas sampling functions such as a gas aspirator and a
dehumidifier are not required.
2. Easy maintenance
The sensor equipped with the detector, has unit construc-
tion, it is easy to replace.
By separating the detector and the flow guide tube, filter
replacement is easy.
3. More reliable than sensor diagnosis, sensor recover-
able function
Depending on the concentration of the measurement gas,
the power of the sensor might deteriorate. The equipment
includes sensor recovery function electronically, checking
the deterioration status of the sensor depletion.
Therefore, it has high reliability and long-lasting stability.
4. Safe and secure
System detects thermocouple break for heater control on
the sensor side. Safety functions of isolating power supply
to the detector or isolating power via external contact input
are also.
5. Easy operation
The operation and setting for the converter can be per-
formed interactively, and available as English, Japanese or
Chinese for language display.
<IP66>
Converter (ZKM1)
<IP67>
Converter (ZKM2)
SPECIFICATIONS
General Specifications
Measuring object: Oxygen in noncombustible gas
Measuring method:
Directly insert type zirconia system
Measuring range: 0 to 2 ··· setting range at option 2 in 50
vol% O2
(in 1 vol% O2 steps)
Repeatability: Within ±0.5%FS
Linearity:
Within ±2%FS
Response time: Within 4 to 7 sec, for 90% (from calibra-
tion gas inlet)
Warmup time: More than 10 min
Analog output: 4 to 20mA DC (allowable load resistance
less than 500W) or 0 to 1V DC (output
resistance more than 100W)
Power supply: Rated voltage;
100 to 120V AC (operating voltage 90
to 132V AC)
200 to 240V AC (operating voltage
190 to 264V AC)
Rated frequency; 50/60Hz
Power consumption:
Maximum 240VA (Detector: approx.
200VA, Converter: approx. 40VA)
Normal 70VA (Detector: approx. 50VA,
Converter: approx. 20VA)
EDSX3-137d
Date Apr. 1, 2011