EDSX3-119B Datasheet, PDF (1/8 Pages) Fuji Electric – IN-SITU ZIRCONIA OXYGEN ANALYZER < ZIRCOMAT-C >
IN-SITU ZIRCONIA OXYGEN ANALYZER < ZIRCOMAT-C >
DATA SHEET
ZFK, ZRY
This oxygen analyzer is used to continuously measure
oxygen concentration in combustion exhaust gas of industrial
boilers or furnaces, and is ideally suited for combustion moni-
toring and control.
The detector (ZFK) used with the analyzer is directly
inserted into the objects measured, eliminating the need for a
sampling device and provides quick response.
The converter (ZRY) features 3 measuring ranges, one
touch calibration and NEMA4 housing.
Comparing to the current converter (ZRM) , this ZRY is
recognized as economical type with simple functions.
General-use detector
FEATURES
1. Sampling device is unnecessary
Gas sampling devices such as a gas aspirator, a dehu-
midifier, etc. are unnecessary because of use of direct-
insertion type detector.
2. High speed response
The adoption of a flow guide tube utilizing the flow of the
measured gas assures quick response (less than 7sec).
3. Dust-tight and water-proof housing
Aluminum die casting housing of converter satisfies
NEMA4 and IP65 standard.
4. 3-measuring ranges and one touch calibration
The converter has 0-5, 10, 25 vol% 3 ranges easily
changed by pin connection and zero and span are simply
calibrated by key operation.
SPECIFICATIONS
General
Measuring object: Oxygen contained in noncombustible
gas
Measuring principle:
Direct-insertion zirconia system
Measuring range: 0 to 5, 10 or 25 vol% O2 (Changed by
internal set pin)
Repeatability:
Within ± 1.0% of full scale
Linearity:
± 2% of full scale
Response time: Within 7sec for 90% response (from
calibration gas inlet)
Power supply: 100, 115, 220 or 230V AC, 50/60Hz
Power consumption:
(approx.)
15 + 50VA (at steady state)
15 + 200VA (at start)
Warmup time: Approx. 15min
Converter
Oxygen detector (ZFK2, 5)
Measuring detector:
For general-use: ZFK2
For corrosive gas: ZFK5
Measured gas temperature:
Flow guide tube system; –20 to +600°C
Measured gas pressure:
–3 to +3kPa
Flow guide tube: Flange; JIS5K 65A FF
(JIS5K-80AFF for high particulate gas)
Insertion length; 0.3, 0.5, 0.75, 1m
(0.8m for high particulate gas)
Ambient temperature:
–20 to +60°C for cable section
125°C or less at detector flange surface
with power applied
Structure:
Dust/rain-proof structure(IEC IP55
equivalent)
Filter:
Alumina(filtering accuracy 50µm) and
quartz paper
Main materials of gas-contacting parts:
General-use detector(ZFK 2); Zirconia,
SUS316, platinum, SUS304
Anticorrosive detector(ZFK 5); Zirconia,
tita-nium, platinum, SUS316
Flow guide tube; SUS304 or SUS316
EDSX3-119b
Date Mer. 6, 2007