English
Language : 

EDSX10-70C Datasheet, PDF (1/8 Pages) Fuji Electric – MICROJET RECORDER-E
MICROJET RECORDER-E
DATA SHEET
PHE
This recorder can record up to 6 channels of thermo-
couples, resistance bulbs and DC voltage/current signals.
The adoption of ink jet system makes it possible to record
measured data in analog trace and digital color printing (6
colors, max) on a 100mm wide chart paper.
FEATURES
1. Compact size
Depth: 175mm, mass; about 1.2kg (continuous type)
Depth: 197mm, mass; about 1.5kg (6-intermittent type)
Ideally suited for use with machines and equipments.
2.High-quality recording
• Continuous traces without pen offset are possible by
our unique ink jet system.
• Scales are printed on a chart paper for each channel,
eliminating the need for scale plate.
• 6 different scale on 6 intermittent recording universal
input type is available.
3.Easy setting of input signals
DC voltage input (5mV span, 50V max.), DC current
(4-20mA, 10-50mA), 12 kinds of thermocouples and re-
sistance bulbs (Pt100) are field-settable for each channel.
4. Digital printing
In addition to analog recording of measured data, peri-
odic data printing,measured value list, scale printing,
alarm printing, burnout printing, and parameter list are
also available.
5.Easy Operation
• A cartridge type recording device is used for easy
replacement.
• Allow to draw out the chart paper while recording.
SPECIFICATIONS
Input system
Input points: 1, 2-continuous recording, 6-intermittent
recording
Input signal: Thermocouple: B, R, S, K, E, J, T, N, W, L,
U, PN
Resistance bulb: Pt100
DC voltage: 5 0 m V, 5 0 0 m V, 5 V, 5 0 V
range
DC current: 4 to 20mA DC, 10 to 50mA
DC (Shunt resistor (option)
need to be connected to
the terminal.)
Input signal setting and change:
Setting and change of input signal be-
tween thermocouple, resistance bulb
and DC voltage (50mV, 500mV, 5V, 50V
range) is possible for each channel by
the setting pin in the instrument and
keyboard operation.
Measuring range (Recording range):
Recording range on each channel is set-
table within the raference range with
keyboard operation.
Measurement cycle:
1, 2-continuous: 0.2s/point
6-intermittent: 30s/all points
Burnout:
When thermocouple or resistance bulb
input is disconnected, the recording is
deflected to 100%.
Input filter:
Settable within the range of 0-255s by
1s. steps.
Initial set before delivery is 3 s.
Reference range
Input signal
Thermocouple B
R
S
K
E
J
T
N
W
L
U
PN
Resistance bulb Pt 100
DC voltage
±50mV
±500mV
±5V
±50V
Scaling
°C
°F
400 to 1760
0 to 1760
0 to 1760
-200 to 1370
-200 to 800
-200 to 1100
-200 to 400
0 to 1300
0 to 1760
-200 to 900
-200 to 400
0 to 1300
752 to 3200
32 to 3200
32 to 3200
-328 to 2498
-328 to 1472
-328 to 2012
-328 to 752
32 to 2372
32 to 3200
-328 to 1652
-328 to 752
32 to 2372
-200 to 600
-328 to 1112
-50.00 to 50.00mV
-500.0 to 500.0mV
-5.000 to 5.000V
-50.00 to 50.00V
Scaling is possible within the range of
-32767 to +32767 (decimal point may
be put as necessary)
Recording system
Writing system: Ink jet system, in 6 colors as max.
Chart width: 100mm
Chart length: Z fold 15.08m
Service life of ink (depends on operating conditions):
About 12 months for 1 continuous line
recording at 20mm/h of chart speed.
EDSX10-70c
Date Apr. 1, 2011