EDS8-49E Datasheet, PDF (1/2 Pages) Fuji Electric – FCX SERIES FIELD INDICATOR
FCX SERIES FIELD INDICATOR
DATA SHEET
FXE
This is a field installation type moving-coil indicator which
receives and indicates signal of 4 to 20 mA DC. It is employed
when an indicator is to be installed separately from the
transmitter.
SPECIFICATIONS
Functional Specifications
Measured input: 4 to 20mA DC (2-wire type)
Internal impedance:
3.5 Ω or less
Scale:
0 to 100% linear scale or 0 to 100%
square scale, actual scale
Note: Double scale is unavailable.
Scale length: 50mm
Environmental protection:
IEC IP67 and NEMA4X
Explosionproof structure:
Intrinsic safety BASEEFA EExiaIICT4,
T5(pending)
Flameproofing JIS d2G4
Case material: Aluminum alloy
Finish color:
Epoxy urethane double coating
Silver (case cover is blue-silver)
Ambient temperature:
– 40 to +85°C
For explosionproof units, ambient temper-
ature must be within the limits specified
by each standard.
Ambient humidity:
100% RH or less
Storage temperature:
– 40 to +90°C
Mass:
Approx. 3.5kg
Conduit connection:
G1/2, 1/2– 14NPT, Pg13.5 or M20 x 1.5
conduit, as specified
External terminal:
Input signal terminal x 2, ground terminal
x 1, M4mm screw
Mounting method:
Mounting on 50A (2B) pipe or on wall
Outer dimensions (H x W x D): 92 x 222 x
179mm
[Performance Specifications]
Accuracy:
± 1.5 class
Withstand voltage:
500V AC, 50/60 Hz, for 1 minute, to ground
Insulation resistance:
100MΩ or more at 500V DC, to ground
SCOPE OF DELIVERY
Indicator and pipe mounting bracket (as specified)
CODE SYMBOLS
1 2 3 4 5 6 7 8 9 10 11 12
FXE 0001 -
Description
Conduit connection
S
G1/2
T
1/2-14NPT
V
Pg13.5
W
M20 x 1.5
Scale
B
0 to 100% linear scale
C
0 to 100% square scale
D
Actual scale
Explosionproof structure
A
General type (non-explosionproof)
B
JIS flameproof (conduit seal type)(*)
C
JIS flameproof (cable seal type)(*)
K
BASEEFA Intrinsic safety
Pipe mounting bracket
A
Not provided
B
Provided (steel)
C
Provided (stainless steel)
Stainless steel tag plate
Y None
B Yes
Note: *10th code B, C is only applicable for 4th
codes.
EDS8-49e
Date Jan. 18, 2003