English
Language : 

EDS6-69B Datasheet, PDF (1/16 Pages) Fuji Electric – INTEGRAL TYPE ELECTROMAGNETIC FLOWMETER
INTEGRAL TYPE ELECTROMAGNETIC
FLOWMETER
DATA SHEET
FMA1, 2
The electromagnetic flowmeter is an instrument to measure
the volumetric flow rate of liquid simply by applying a mag-
netic field from the outside utilizing the fact that an electric
conductor which crosses a magnetic field at a certain velocity
causes voltage to be induced in proportion to the velocity,
which is known as Faraday’s law.
This flowmeter is designed with the latest electronics tech-
nology and magnetic/electric field analyzing technology, real-
izing a compact and light-weight structure and measurement
with high accuracy.
FEATURES
1. High accuracy
The adoption of the optimum magnetic field design using
the 3-dimentional finite element method has minimized
the effects of flow velocity profile and materials of adja-
cent piping. At flow velocity of more than 1m/sFS,the
measurement accuracy is as high as ± 0.5% of
indicatedvalue when the flow is above 20%FS.
2. Wide range
Measurement range: 0 to 0.1 ... 15m/sec in flow velocity
3. High-reliability structure
The converter case is a sealed 2-chamber structure, practi-
cally free from dew condensation and sudden submer-
gence. The adoption of stainless housing assures excel-
lent anti-environment efficiency.
4. Easy-to-see display
LCD with back-light allows easy check of display even in a
dark place. Instantaneous flow and integral volmetric flow
are displayed at the same time.
5. Free power supply
The flowmeter operates on power supply 100 to 240V AC,
50/60 Hz, eliminating the need for selection of power volt-
age.
6. Application of international standards
The overall length of flange type flowmeter conforms with
ISO draft standards. (Meter size : 6A-200A)
SPECIFICATIONS
Sensor
• Measurement item :
General-use industrial water, tap-water,
sewage, waste water, chemicals slurry,
and other liquids with conductivity of more
than 3µs/cm.
• Structure : Wafer type or flange type
• Mounting method :
Mounted via flange insertion type on oppo-
site (wafer type) (with Guide rings)
··· 2.5 to 200A
Note 1) Guide ring : A ring-shaped guide
used for centering the sensor when a wa-
ter type is mounted on the piping.
or flange mounting
··· 6 to 300A
Note 2) Flange with meter size 2.5A or 6A
can be used for 15A.
• Liquid pressure :
-100 to 2000kPa or flange operating
• Meter size and measurement range
Measuring range is equivalent to flow ve-
locity 0.1 to 15m/s.
Meter size
Min.measurement range Max. measurement range
[m3/h]
[m3/h]
2.5A Note 3)
6A
15A
25A
40A
50A
80A
100A
150A
200A
250A
300A
0 to 0.00177
0 to 0.0102
0 to 0.0637
0 to 0.177
0 to 0.453
0 to 0.707
0 to 1.81
0 to 2.83
0 to 6.37
0 to 11.3
0 to 17.7
0 to 25.5
0 to 0.265
0 to 1.52
0 to 9.54
0 to 26.5
0 to 67.8
0 to 106
0 to 271
0 to 424
0 to 954
0 to 1696
0 to 2650
0 to 3817
Note) Meter size 2.5A: Wafer type only
EDS6-69b
Date Oct. 29, 2001