EDS3-139B Datasheet, PDF (1/16 Pages) Fuji Electric – IN-SITU ZIRCONIA OXYGEN ANALYZER FOR HAZARDOUS LOCATION
IN-SITU ZIRCONIA OXYGEN ANALYZER
FOR HAZARDOUS LOCATION
DATA SHEET
ZFKE, ZKME
This oxygen analyzer is used to continuously measure
oxygen concentration in noncombustible exhaust gas of
industrial boilers or furnaces, and is ideally suited for com-
bustion management and control.
The analyzer system is comprised of the detector and
converter coupled together as a complete system. Detec-
tor setting configuration includes the flow guide tube and
detector sensor. The flow guide tube is inserted directly
into the gas and directs gas to the sensor for measurement.
The converter (ZKME) is comprised of the signal proces-
sor, input/output and communications, display and system
controls.
The converter is equipped with advanced functional-
ity such as performing the sensor diagnostics and sensor
recovery function, so the detector can be used within long
term stability.
Detector (ZFKE)
FEATURES
1. Gas sampling device is unnecessary
For quick response, insert the detector directly into the
staek. Gas sampling functions such as a gas aspirator
and a dehumidifier are not required.
2. Easy maintenance
The sensor equipped with the detector, has unit construc-
tion, it is easy to replace.
By separating the detector and the flow guide tube, filter
replacement is easy.
3. More reliable than sensor diagnosis, sensor recoverable
function
Depending on the components in the measurement
gas, the characteristics of the sensor might deteriorate.
The equipment includes sensor recovery function elec-
tronically, checking the deterioration status of the sensor
depletion.
Therefore, it has high reliability and long-lasting stability.
4. Safe and secure
System detects thermocouple break for heater control on
the sensor side. Safety functions of isolating power sup-
ply to the detector or isoralting power via external contact
input are also.
5. Easy operation
The operation and setting for the converter can be per-
formed interactively, and available as English, Japanese
or Chinese for language display.
Converter (ZKME)
SPECIFICATIONS
General Specifications
Measuring object: Oxygen in noncombustible gas
Measuring method:
Directly insert type zirconia system
Measuring range: 0 to 2 ··· 0 to 50 vol% O2
2 ranges available in 1 vol% O2 steps
Repeatability:
Within ±0.5%FS
Linearity:
Within ±2%FS
Response time: Within 4 to 7 sec, for 90% (from calibra-
tion gas inlet)
Warmup time: approx. 10 min
Analog output: 4 to 20mA DC (allowable load resis-
tance less than 500Ω) or 0 to 1V DC
(output resistance more than 100Ω)
Power supply: Rated voltage;
100 to 120V AC (operating voltage 90
to 132V AC)
200 to 240V AC (operating voltage
190 to 264V AC)
Rated frequency; 50/60Hz
Power consumption:
Maximum 240VA (Detector: approx.
200VA, Converter: approx. 40VA)
Normal 70VA (Detector: approx. 50VA,
Converter: approx. 20VA
EDS3-139b
Date Apr. 1, 2011