COMPACT TYPE GAS ANALYZER
DATA SHEET
ZSVS
OVERVIEW
The compact type gas analyzer with built-in pump and filter
is intended for heat treat furnace, plant cultivation, and re-
search-purpose chemical analysis. With the gas extractor,
either simplified measurement probe (non-fixed type) or con-
tinuous measurement probe (fixed type) is selectable at
option. Since a high-sensitivity single-beam mass flow con-
troller is adopted for the infrared sensor, long-term stability
and maintainability are excellent.
FEATURES
1. Gas concentrations of 4 components measurable simul-
taneously and continuously (Note).Concentrations of max.
3 gas components among CO2, CO and CH4 are continu-
ously measurable with the infrared sensor, and that of
O2 is continuously measurable with galvanic cell type
oxygen sensor.
2. Standard with 3-range analyzer
Measurement can be performed over a broad range us-
ing the range selector key.
3. Compact type small and easy to use.
The compact body with built-in pump and membrane filter
is easy to carry and install.
Operation is easier because operation keys and display
unit are all gathered in the analyzing block.
4. A rich variety of standard functions incorporated.
Max. 8-channel outputs are allowed including instanta-
neous concentration value, O2-corrected value, O2-cor-
rected moving average value and CP calculation value
outputs.
(Note) For continuous measurement, the standard require-
ments for sample gas (shown on page 3) need to be
met, and zero/span calibrations and membrane filter
replacement are required periodically.
SPECIFICATIONS
Standard Specifications
• Measuring system:
CO2, CO and CH4; Non-dispersive infra-
red absorption method with single light
source and single beam (single beam
method)
O2; Galvanic cell method
• Measurable component and min./max. measuring range:
CO2; 0 to 200 ppm / 0 to 100 %
CO; 0 to 200 ppm / 0 to 100 %
CH4; 0 to 1000 ppm / 0 to 100 %
O2; 0 to 5 % / 0 to 25 %
Max. 4 components measurable includ-
ing O2
• Number of measuring ranges:
3 ranges
• Max. range ratio 1:5
• Warm-up time: 30 min after power-on
Provided with count-down timer indicat-
ing function.
• Analog output: In up to 8 channels.
4 to 20 mA DC or 0 to 1 V DC (linear)
Non-isolated output
Allowable load; 4 to 20 mA DC, 550Ω or
less
0 to 1 V DC, 100 kΩ or
more
• Instantaneous value output of each gas
component
• Instantaneous value output after O2 cor-
rection (when provided with O2 analyzer)
• Average value output after O2 correc-
tion (when provided with O2 analyzer)
• CP calculation value output (when pro-
vided with CO2 analyzer)
* The channel numbers of indicated value
and output value correspond to each
other one by one.
*An exclusive 25-pin cable is standard-
equipped.
• Communication output:
RS-232C Modbus protocol
*Use a commercially available product
(D-sub 9-pin cable).
This product is not explosion-proof. When handling dangerous gas, adequate attention shall be paid.