EDS3-128 Datasheet, PDF (1/8 Pages) Fuji Electric – CO/O2 GAS ANALYZER FOR STACK GAS | |||
|
CO/O2 GAS ANALYZER FOR STACK GAS
DATA SHEET
This instrument is equipped with a non-dispersive infrared
CO sensor and a galvanic oxygen sensor and used for
measuring CO and O2 in the gas emitted from garbage in-
cinerators.
CO instantaneous value, O2 instantaneous value and CO
average value after O2 correction (1-hour and time-variable
averaging) can be output.
ZSW
FEATURES
1. The CO sensor adopts high-performance and long-life non-
dispersive infrared ray (NDIR) system. The CO sensor is
not consumable, and need not be replaced in 1 or 2 years.
2. Automatic calibration function by air equipped as stan-
dard allows stable measurement. No need to worry about
the consumption of standard gas.
3. Space-saving configuration unitized into 19-inch type al-
lows the maintenance from the front.
4. A microjet recorder (Fuji Electricâs type PHC or PHE) can
be built in for recording measured gas concentration and
temperature values.
SPECIFICATIONS
⢠Measuring system:
Non-dispersive infrared absorption (NDIR)
method for CO
Galvanic method for O2
⢠Measurable component and range:
CO ; 0~500/1000ppm
0~500/2000ppm
O2 ; 0~25vol%
⢠Repeatability: ± 0.5% of full scale/CO
± 1% of full scale/O2
⢠Zero drift:
Max. ± 2.0% of full scale/month for CO
Max. ± 2.0% of full scale/week for O2
⢠Span drift:
Max. ± 2.0% of full scale/week for CO
Max. ± 2.0% of full scale/month for O2
⢠Linearity:
Max. ± 2.0% of full scale
⢠Sample gas extracting rate:
Approx. 0.4R/min
⢠Response time: Within 150 seconds for 90% indication
(after extracting sample gas through the
inlet)
⢠Warm-up time: Within 4 hours after power-on
⢠Output signals: Each signal within range from 4 to 20 mA
DC
Non-isolated (isolated output available at
option)
CO instantaneous value
O2 instantaneous value
Moving average CO value after O2 cor-
rection (averaging time variable)
Time setting in 1 to 59 minutes (1-minute
increment) or 1 to 4 hours (1-hour incre-
ment)
Allowable load resistance 550 ⦠or less
⢠Pump ON/OFF contact input:
No voltage contact (Closed: Pump ON)
⢠Indication:
LCD with back light for indicating CO in-
stantaneous value, O2 instantaneous
value, CO instantaneous value after O2
correction and CO average values after
O2 correction, O2 average value (1-hour
and time-variable averaging) and param-
eter assignment
⢠Recorder (option):
100 mm-width recorder (Fuji Electricâs
type PHC or PHE) built in, max. 6 points
recordable
⢠Gas extractor: General type (without filter)
⢠Titanium probe (300, 400, 600 or 800
mm long)
⢠Flange JIS 5K25AFF
⢠Weight: Approx. 1 kg (800 mm)
Electrical heating type (filter built in)
⢠Filter mesh; 40µm mesh of SUS 316
stainless steel
⢠Probe; SUS 316 stainless steel of 300,
400, 600 or 800 mm length
⢠Flange; JIS 5K65AFF
⢠Weight; Approx. 8.5 kg
EDS3-128
Date Feb. 20, 2004
|
▷ |