EDS3-117 Datasheet, PDF (1/2 Pages) Fuji Electric – PORTABLE GAS ANALYZER
PORTABLE GAS ANALYZER
DATA SHEET
This unit is a simple and stable analyzer using a single-beam
infrared-ray system. It is designed for measuring carbon
dioxide gas (CO2), carbon monoxide (CO) or methane (CH4),
and is used to measure atmospheric gases in agricultural
vinyl greenhouse, carbon dioxide gas in rooms or carbon
dioxide gas, carbon monoxide, ethane, etc., in thermal pro-
cessing furnaces.
ZFY
FEATURES
1. Single-beam NDIR system is used for simple structure,
providing excellent stability without interference of other
gases.
2. The adoption of a unique mass-flow type sensor assures
high sensitivity and long service life.
3. One-touch zero/span calibration, self-diagnosis function
and 2-range select function assure easy operation and
maintenance.
4. Sampling devices such as pump, filter, flow checker, etc.,
are built-in to allow the user to make measurements sim-
ply by connecting a sampling tube.
5. Compact design offers easy carrying and handling.
SPECIFICATIONS
1) Application : Facilities gardening, measurement of
gases in rooms, combustible exhaust gas,
gases in thermal processing furnaces or
ceramic industry furnaces, and various
tests
2) Measurement principle :
NDIR (Non-dispersive Infrared-absorption
method; single-beam system)
3) Measuring components and measurement range :
Measurement
range
0 to 0.2/0.5%
0 to 0.5/1.0%
Measuring
components CO2 CO CH4
™× ×
™™ ×
0 to 1.0/2.0%
™™™
0 to 2.0/5.0%
™™™
0 to 5.0/10.0%
0 to 10.0/20.0%
0 to 20.0/50.0%
0 to 50.0/100.0%
™™™
™™ ×
™™ ×
™× ×
4) Repeatability : ± 0.5%FS
5) Drift :
± 2%FS /day for zero and span
6) Linearity : ± 2%FS
7) Response : 90%, less than 15sec
8) Sample flow : About 1R/min (pump, filter and flowme-
ter, built-in)
9) Analog output :
4 to 20mA DC (allowable load; less than
550Ω)
10)Indicator : 4th code digital display
11)Calibration : One-touch calibration using canned gas
or compressed standard gas (φ6 tube
connection attachment supplied)
12)Warming time :
About 30min
13)Gas connecting parts material :
SUS, polyethylene, polypropylene, CaF2,
toalon, chloroprene
14)Power source :
100V or 115V AC ±10%, 50/60Hz, about
40VA (power cord supplied)
15)Self-diagnosis function :
Calibration range over, analyzer error, etc.
16)Sample gas condition :
Dust; 0 to 10mg/Nm3(filter should be used
at the front for continuous measurement)
Temperature; Less than 50˚C at analyzer
inlet Pressure; –2.94 to +2.94kPa {–300
to +300mmH2O}
17)Installation condition :
Temperature; 0˚C to 40˚C
Humidity; 90%RH or less
Installation place; indoor (Do not install
in a place near combustible exhaust gas.
Be sure to install on a flat stand.)
18)Case :
Plastic, portable type (with ventilating
holes)
19)Dimensions : (H)170 × (W)260 × (D)375mm
20)Mass :
Approx. 5kg
EDS3-117
Date Dec. 18, 1998