EDS3-110D Datasheet, PDF (1/16 Pages) Fuji Electric – NDIR TYPE INFRARED GAS ANALYZER
NDIR TYPE INFRARED GAS ANALYZER
DATA SHEET
ZRG
This NDIR gas analyzer features high accuracy, multiple
functions and easy operation through use of a microproces-
sor. It also utilizes a mass flow detector noted for its high
sensitivity and reliability.
This analyzer is used not only for measuring environmen-
tal pollution but is also used for various processes and/or
experiments.
This product, combined with sampling system (ZSP), has
passed the type-approval test based on Weight and Mea-
sure Act.
FEATURES
(1) Use of a microprocessor provides high accuracy, mul-
tiple functions and easy operation.
• A maximum of 3 gas components can be calibrated
with the built-in automatic calibrator (option).
• Using a signal from a zirconiz O sensor (ZFK3) or from
2
any other O2 sensor, the gas analyzer is able to out-
put a value converted into O2 (option).
• Alarm function to emit high/low limit contact output
is also available (option).
• Measuring range can be selected by using external
single (option).
• Zero and span can be calibrated with high accuracy,
simply by pressing calibration keys.
• A self-diagnosis function is included.
(2) This analyzer utilizes mass flow detector featuring high
sensitivity and reliability. It has 2 standard measuring
ranges with a range ratio of up to 1:20.
(3) Three different gas components can be measured si-
multaneously by connecting a zirconia O2 sensor (ZFK3),
in addition to one-and two-component type sensors.
(4) Besides the standard measurement type, a sample swi-
tching type and a differential flow type are also available.
SPECIFICATIONS
General items
Power supply: 100V ± 10% AC, 50/60Hz
115, 220V AC also available on request
Note: Refer to "Code symbols"
Power consumption:
125VA max.
Ambient temperature:
–5 to +45°C
Ambient humidity:
90% RH or less
Enclosure:
Steel casing, indoor-use
Dimensions (H x W x D):
835 x 220 x 232 mm
Mass{weight}: Approx. 24kg
Finish color: Munsell 2.5Y 8.4/1.2
Indication:
4-digit LED for concentration
4-digit LED for sub-indication
Output hold: Output hold prior to manual/auto calibra-
tion is possible. Hold ON/OFF is select-
able.
Standard requirements for sample gas:
Temperature 0 to 50°C
(dehumidification 2°C satu-
ration or less)
Dust
0.3µm or less
Pressure 9.8kPa{0.1kgf/cm2} or less
(flow rate 0.5R/min)
Standard control for sample gas:
Calibration gas Dry gas
Interfere control gas 2°C saturation
Warm-up time: About 8 hours (after power ON)
About 4 hours for sample switching type
Material of gas-contacting parts:
Sample cell; 304 stainless steel, neoprene
rubber
Infrared-ray transmitting window; CaF2 or
sapphire
Internal tubing; Teflon tube, silicone tube,
toaron tube
Gas inlet/outlet, purge gas inlet size:
Rc 1/4 (PT 1/4 internal thread) or NPT1/4
internal thread
EDS3-110d
Date Jun. 29, 2005