EDS3-109E Datasheet, PDF (1/8 Pages) Fuji Electric – INFRARED GAS ANALYZER FOR STACK GAS
INFRARED GAS ANALYZER FOR STACK GAS
DATA SHEET
ZSP
This instrument consists of an (or two) infrared gas analyzer(s),
a zirconia O2 sensor and a sampling device. It is used for
measuring the NOx, SO , CO and O components in the flue
2
2
gas of various boilers, garbage incinerators, furnaces, etc.
through a selective combination of the components.
The instrument has passed a type approval test in compliance
with regulations on measurement and weighing (specified in
Japan) of each component. It is extensively applicable as a
gas analyzer for environmental pollution control and com-
bustion control.
FEATURES
1. Simultaneous measurement of five components
A combination of two infrared gas analyzers plus a zirconia
02 sensor allows simultaneous measurement of four
components, NOx, SO2, CO, CO2 and O2.
2. Easy operation with multiple functions
This analyzer can be equipped with a number of functions
such as auto calibration, O2 correction and self-diagnosis.
It is easier to operate and maintain than a separate type
calculator system.
3. Little interference from other gases
Measuring error due to interfering gases contained in
sample is nearly zeroed by the Permapure dryer and
interference compensating detector adopted in the ana-
lyzer section respectively.
4. Excellent measuring accuracy and stability
The optical system is so constructed as to be free from
external stress by using a single light source and double
beam paths, and therefore assures improved measuring
accuracy plus excellent stability.
5. High sensitivity and wide dynamic range
The infrared gas analyzer, which mounts a mass flow
sensor, has a high sensitivity and a broad dynamic range,
and is capable of switching the measuring range to obtain
1:20 at maximum.
6. Reliable sampling system
Simple but reliable sampling system based on Fuji's rich
experience in the field enables easy maintenance and
reduces frequency of maintenance.
SPECIFICATIONS
Type of cubicle: Indoor type or outdoor installation type
Measuring system:
Non-dispersion infrared absorption
(NDIR) method for NOx, SO2 and C0;
zirconia method for O2
Sampling system: Dry sampling at approx. –15°C (dew
point) (Permapure dryer adopted)
Measuring range:NOx: 0 to 100…2000ppm
SO2: 0 to 100…1000ppm
CO: 0 to 200…2000ppm
CO2: 0 to 5…50%
O2: 0 to 10/25%
Dual range (except forO2), max. range
ratio 1:20
Note: A type test-satisfied instrument having a measuring range of 50
to 2000ppm is available for each of NOx, SO2 and CO.
Repeatability: ± 0.5% of full scale
Drift:
Zero; ± 2% of full scale/week for NOx,
SO2 and CO
± 2% of full scale/month for O2
Span; ±2% of full scale/week
Linearity:
± 2% of full scale
Response time: Within 3 minutes for 90% indication (after
extracting sample gas through the inlet)
Output:
Within 4 minutes for SO2
4 to 20mA DC (allowable load resistance
550Ω), non-isolated
For the number of output points, see
Table 1.
Contact output: Each NO (1a) contact (100V AC, 1 A) for
error (analyzer error, calibration error) and
auto calibration status
Indication:
Digital indication (on analyzer front panel)
Sample gas extracting rate:
Approx. 2L/min
Gas extractor: Electrical heating type (filter built in)
Filter mesh;40µm mesh of 316 stainless
steel
Probe;
316 stainless steel, length
600, 800 or l000mm
Flange; JIS 5K65A FF
EDS3-109e