EDS3-108E Datasheet, PDF (1/16 Pages) Fuji Electric – MICROPROCESSOR BASED NDIR GAS ANALYZER
MICROPROCESSOR BASED
NDIR GAS ANALYZER
DATA SHEET
ZRF
This NDIR gas analyzer features a high accuracy, multiple
functions and easy operation through use of a microprocessor.
It also utilizes a mass flow detector noted for its high sensitivity
and reliability. Being housed in a 19 inch rack case suitable
for mounting on a panel or a table top, this analyzer is applicable
not only for measurement of environmental pollution but for
use in various processes and/or experimental laboratories.
FEATURES
(1) Use of a microprocessor provides high accuracy,
multiple functions and easy operation.
• The built-in automatic calibrating function allows
calibration of up to three components (option).
• The signal from a zirconia O2 sensor (ZFK3) or
other O2 meter enables output of an O2 correction
value (option).
• Includes an alarm function providing an upper/
lower limit contact output (option).
• Range can be changed over by external signal
(option).
• Zero and span calibration is accurate and easy by
means of operating keys.
• A self-diagnosis function is included.
RS232C port available (option).
(2) This analyzer utilizes a mass flow detector featuring
high sensitivity and reliability. It is equipped with
two standard ranges for a range ratio of up to 1:20.
(3) Addition of a zirconia O2 sensor (ZFK3) to the one/
two-component analyzer allows measurement of up
to three components simultaneously.
(4) Besides the standard method of measurement, a
sample switching system and differential flow
system are also available.
SPECIFICATIONS
General items
Power supply: 100, 115 or 220V AC ± 10%, 50/60Hz
Power consumption:
125VA max. (220VA max. when CO/CO
2
converter equipped)
Ambient temperature:
–5 to +45°C
Ambient humidity:
90%RH or less
Enclosure:
Steel casing, for indoor application
Outer dimensions (H x W x D):
Rack mounting type;
220 x 483 x 463mm
Panel flush mounting type;
220 x 443 x 463mm
Table top type;
232 x 443 x 463mm
Mass {weight}: Approx. 20kg
Finish color: Munsell 2.5Y8.4/1.2
Indication:
4 digit LED for concentration
4 digit LED for sub-indication
Output hold: Output value before manual or automatic
calibration is held. Whether or not to
effect hold function can be selected.
Sample gas condition:
Temperature; 0 to 50°C (due point of
water vapor; less than 2°C)
Dust; less than 0.3µm
Pressure; less than 9.8kPa
Standard adjustment:
Calibration gas; Dry N Balance
2
Interface compensation Dew point of 2°C
water vapor in N
2
Warm up time: Approx. 4 hours
Material of gas-contacting parts:
Sample cell; SUS304, neoprene rubber
Infrared-ray transmitting window: CaF
2
or sapphire
Internal tubing; Teflon tube, silicone tube,
Toaron tube
Gas inlet/outlet, purge gas inlet size:
Rc1/4 (PT1/4 internal thread) or NPT1/4
internal thread
Purge gas flow rate:
1± 0.5r/min
It is necessary to purge the instrument
interior when ambient air contains the
corrosive gas etc. or the measuring range
of CO is less than 0 to 50ppm.
2
Scope of delivery: Analyzer, mounting bracket, test
report, power fuse, cloth for cleaning
infrared-ray transmitting window
EDS3-108e
Date Jun. 29, 2005