OVERVIEW
This is a paperless recorder that displays measured data on
the LCD in real time and stores data in CompactFlash.
The type of input such as thermocouple, resistance bulb,
D.C. voltage (current), etc. can be arbitrarily set to 36 chan-
nels at the maximum.
The data stored in CompactFlash can be regenerated on the
screen, and the use of supplied support software allows the
data to be regenerated on a PC screen.
The data recorded in ASCII format can be directly read in a
spreadsheet such as Excel, which facilitates the process-
ing on a PC. (The data recorded in binary format cannot
be read in.)
FEATURES
1. Large capacity storage by CompactFlash
Measured data is periodically stored in CompactFlash.
In case of 512 MB, for example, display files for about 2
year and a half (display refresh cycle 30 sec) can be taken
up (in case of ASCII data format, 9 channels, maximum/
minimum recording).
2. Quick search and display of past data
Data stored in CompactFlash can be displayed in succes-
sion by scrolling the screen.
3. Various display capability
Depending on the object of measurement, the most
suitable display format can be selected from a variety of
formats including bar graph display, trend display, digital
display, etc.
4. PC support software supplied as standard
Loader software that enables easy display and change of
set data and data viewer software that regenerates the
data stored in CompactFlash are supplied as standard.
5. 36-point recording
12 types of thermocouples, 2 types of resistance bulbs
and DC voltage/current input can be recorded up to 36
points.
6. LCD extinguishing function
Automatically extinguishes the LCD if nothing is operated
for certain time. You can set the time after a lapse of
which the LCD is extinguished via parameter "LCD ex-
tinguishing time". The settable range is 0 to 60 minutes.
Setting at 0 minute overrides the function, whereby the
LCD will never extinguish.
This function prevents the backlight life from shortening
uselessly. During the extinguishment, the power con-
sumption can be reduced.
7. Ethernet function (Option)
FTP, Web server, e-mail and MODBUS-TCP are available
using 10Base-T.
SPECIFICATIONS
Input system
Number of input points:
9, 18, 27 or 36 points (Can be selected
at the time of purchase)
Input circuit: Input mutual isolation (See ”Others” on
page 4 for the withstand voltage)
Resistance bulb measured current:
about. 1 mA
Measuring cycles:
9 or 18 points….100ms cycles
27 or 36 points….200ms cycles
Recording cycle: 1 second to 12 hours
Input types: Thermocouple, resistance bulb, dc
voltage, and dc current (Shunt resistors
are fitted in input terminals).
Note) Provide a shunt resistor (type:
PHZP0101) separately.
Measuring range
Input types
Reference range
Thermocouple B
R
S
K
E
J
T
N
W
L
U
PN
Resistance bulb JPt100
Pt100
Ni100
Pt50
Cu50
400.0 to 1760.0°C
0.0 to 1760.0°C
0.0 to 1760.0°C
–200.0 to 1370.0°C
–200.0 to 800.0°C
–200.0 to 1100.0°C
–200.0 to 400.0°C
0.0 to 1300.0°C
0.0 to 1760.0°C
–200.0 to 900.0°C
–200.0 to 400.0°C
0.0 to 1300.0°C
–200.0 to 600.0°C
–200.0 to 600.0°C
–60.0 to 180.0°C
–200.0 to 600.0°C
–50.0 to 200.0°C
DC voltage
50mV
500mV
1-5V
0-5V
0.00
0.0
1.000
0.000
to 50.00mV
to 500.0mV
to 5.000V
to 5.000V
Note) B, R, S, K, E, J, T, N :
JIS C 1602, DIN IEC 584-1
W : 5%Re-26%Re · W
(Hoskins Mfg. Co. USA)
L : Fe-Cu · Ni (DIN 43710)
U : Cu-Cu · Ni (DIN 43710)
PN: Platinel
JPt100 : JIS C 1604-1989 (Old JIS Pt 100)
Pt100, Pt50 : JIS 1604, DIN IEC 751
EDS10-79c