English
Language : 

EDS10-74J Datasheet, PDF (1/12 Pages) Fuji Electric – PAPERLESS RECORDER
PAPERLESS RECORDER
DATA SHEET
PHR
This is a paperless recorder that displays measured data
on the LCD in real time and stores data in CompactFlash.
The type of input such as thermocouple, resistance
bulb, D.C. voltage (current), etc. can be arbitrarily set to 18
channels at the maximum.
The data stored in CompactFlash can be regenerated on
the screen, and the use of supplied support software al-
lows the data to be regenerated on a PC screen.
The data recorded in ASCII format can be directly read in
a spreadsheet such as Excel, which facilitates the process-
ing on a PC. (The data recorded in binary format cannot be
read in.)
FEATURES
1. Large capacity storage by CompactFlash
Measured data is periodically stored in CompactFlash.
In case of 512 MB, for example, display files for about
3 year and a half (display refresh cycle 30 sec) can be
taken up (in case of ASCII data format, 9 channels, maxi-
mum/minimum recording).
2. Quick search and display of past data
Data stored in CompactFlash can be displayed in suc-
cession by scrolling the screen.
3. Various display capability
Depending on the object of measurement, the most
suitable display format can be selected from a variety
of formats including bar graph display, trend display,
digital display, etc.
4. PC support software supplied as standard
Loader software that enables easy display and change of
set data and data viewer software that regenerates the
data stored in CompactFlash are supplied as standard.
5. Compact size
160 (W) × 144 (H) × 185 (D) mm(Panel mounting), Com-
pact and as light as about 1.5 kg (9-point input, without
option).
6. 18-point recording (Option)
12 types of thermocouples, 2 types of resistance bulbs
and DC voltage/current input can be recorded up to 18
points
7. Communication function (Option)
RS485 MODBUS communication is available.
8. LCD extinguishing function
Automatically extinguishes the LCD if nothing is oper-
ated for certain time.You can set the time after a lapse
of which the LCD is extinguished via parameter "LCD ex-
tinguishing time".The settable range is 0 to 60 minutes.
Setting at 0 minute overrides the function, whereby the
LCD will never extinguish.
This function prevents the backlight life from shorten-
ing uselessly.During the extinguishment, the power
consumption can be reduced.
9. Ethernet function (Option)
FTP, Web server, e-mail and MODBUS-TCP are available
using 10Base-T.
SPECIFICATIONS
Input system
Number of input points:9 points or 18 points (Can be
selected at the time of purchase)
Input circuit: Input mutual isolation (See "Others" on
page 5 for the withstand voltage)
Resistance bulb measured current:
about. 1 mA
Measuring cycles:9 or 18 points….100ms cycles
Recording cycle: 1 second to 12 hours
Input types: Thermocouple, resistance bulb, dc volt-
age, and dc current (Shunt resistors are
fitted in input terminals).
Note) Provide a shunt resistor (type:
PHZP0101) separately.
Measuring range
Input types
Reference range
Thermocouple B
R
S
K
E
J
T
N
W
L
U
PN
400.0 to 1760.0ºC
0.0 to 1760.0ºC
0.0 to 1760.0ºC
–200.0 to 1370.0ºC
–200.0 to 800.0ºC
–200.0 to 1100.0ºC
–200.0 to 400.0ºC
0.0 to 1300.0ºC
0.0 to 1760.0ºC
–200.0 to 900.0ºC
–200.0 to 400.0ºC
0.0 to 1300.0ºC
Resistance bulb JPt100
Pt100
–200.0 to 600.0ºC
–200.0 to 600.0ºC
DC voltage
50mV
500mV
1-5V
0-5V
0.00
0.0
1.000
0.000
to 50.00mV
to 500.0mV
to 5.000V
to 5.000V
Note) B, R, S, K, E, J, T : JIS C 1602, DIN IEC 584-1
N : NICOSIL-NISIL (IEC 584)
W : 5%Re-26%Re · W (Hoskins Mfg. Co. USA)
L : Fe-Cu · Ni (DIN 43710)
U : Cu-Cu · Ni (DIN 43710)
PN: Platinel
JPt100 : JIS C 1604-1989 (Old JIS Pt 100)
Pt100 : JIS C 1604, DIN IEC 751
EDS10-74j
Date Dec. 5, 2011