English
Language : 

FSP400-1K50M1_15 Datasheet, PDF (1/1 Pages) FSP TECHNOLOGY INC. – AC -DC switching power supplies
Medical PSU
FSP400-1K50M1
DE SC RIP TIO N
This series of AC -DC switching power supplies in a
package of 4 x 7 x 1.58 inches are capable of delivering
400 watts of continuous power at 7 C FM forced air cooling
or 300 watts at convection cooling. The units are
constructed on a printed circuit board with a U-bracket for
mechanical support and heat sinking. A cover and fan
assembly can be added during manufacturing for 400 watt
output without the change of any dimension. They are
designed for medical applications, but not for life-
supporting equipment. The units are certified also to
IEC /EN/UL 60950-1 and suitable for data networking,
computer and telecommunication applications.
SAFETY STANDARD APPAOVAL
FEATURES
4 x 7 inch footprint with 1.58 inch low profile
100-240 VAC input with active PFC
Less than 300 μA leakage current
300 watt convect ion rating up to +50℃
400 watt output with 7 C FM forced air
Standby output 5VDC a t 100mA
EN55011 / 5 5022 C lass B conducted emissions
Inhibit - TTL low to disable output
Standard PS Off and DC OK signals
Efficiency greater than 88%
C ompliant wi th RoHS requirements
W ATTAGE
Wa tta g e :
400W
OUTPUT SPECIFICATION
Ripple & Noise:
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500 us after a
25% step load change
Over Current
Protected to output short
Protection:
circuit conditions
ENVIRONMENTAL SPECIFICATION
T E MP.Ra ng e :
Operating Temperature:-10°C to
+70°C
Storage Temperature: -40°C to +
85℃
DIMENSION
Dimension:
177.8mm(L) x 101.6mm(W) x
39.6mm(H)
INPUT SPECIFICATION
Input Range:
Input Frequency:
Input Current:
Leakage Current:
90-264 Vdc
47-63 Hz
4.2A(rms) for115VAC ,
2.1A(rms) for230VAC
300 µA max. @ 264 VAC ,63
Hz
*Output Voltage and C urrent Rating
Ripple-Noise(R-P) mV
Regulation Load %
Output Max.(A)
Output Min.(A)
+28V
280mV
±2%
14.29A
0A
NO TE S
1. C hange suffix “B” for U-Bracket form to “C ” for enclosed form with cover and fan assembly, e.g. PM400-14C .
2. 300 W without moving air or 400 W with 7 C FM forced air provided by user for “B” version, 400 W for “C ” version with
cover and fan assembly
3. Ripple and noise is maximum peak to peak voltage value measured at output within 20 MHz bandwidth, at rated line
voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the
output.
MECHANICAL SPECIFICATION
This content is subject to change, please refer to specification for m ore detail.
FSP reserve the right to change the content without prior notice