English
Language : 

FSP180-AHAM1 Datasheet, PDF (1/1 Pages) FSP TECHNOLOGY INC. – Medical PSU
Medical PSU
FSP180-AHAM1
DE SC RIP TIO N
This series of AC /DC switching power supplies are for 180
watts of continuous output power. They are enclosed in a
94V-1 rated polyphenylene-oxide case with an inlet of the
IEC 320/C 14 or IEC 320/C 8 to mate with interchangeable
cord for world-wide use. All models meet EN 55011 and
FC C class B emission limits, and are designed for medical
applications not for life-supporting equipment.
FEATURES
High efficiency
Low ripple & noise
Over voltage protection
Short-circuit protection
Overpower protection
Over temperature protection
100% burn-in at full rated load
Standby consumption less than 0.5 W
C ompliant with C EC and ENERGY STAR efficiency level V
requirements
C ompliant with RoHS requirements
W ATTAGE
Wa tta g e :
180W
SAFETY STANDARD APPAOVAL
OUTPUT SPECIFICATION
Ripple & Noise:
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500 us after a
25% step load change
Over Current
All models protected to short-
Protection:
circuit conditions
DIMENSION
Dimension:
INPUT SPECIFICATION
Input Range:
Input Frequency:
Input Current:
Leakage Current:
197.0mm(L) x 88.0mm(W) x
44.0mm(H)
90-264 Vdc
50-60 Hz
2.4A(rms) for115VAC ,
1.2A(rms) for230VAC
200 µA max. @ 264 VAC ,63
Hz
ENVIRONMENTAL SPECIFICATION
T E MP.Ra ng e :
Operating Temperature:0°C to
+40°C
Storage Temperature: -20°C to +
80℃
OUTPUT ELECTRICAL
Single Output:
12 V/ 15 A
NO TE S
1. C lass I models are equipped with IEC 320/C 14 inlet, and class II models with IEC 320/C 8 inlet.
2. Ripple and noise is maximum peak to peak voltage value measured at output within 20 MHz bandwidth, at rated line
voltage and output load ranges, and with a 47 μF electrolytic capacitor in parallel with a 0.1 μF ceramic capacitor across
the output.
This content is subject to change, please refer to specification for m ore detail.
FSP reserve the right to change the content without prior notice