English
Language : 

FSP105-KLAM1 Datasheet, PDF (1/1 Pages) FSP TECHNOLOGY INC. – Low safety ground leakage current
Medical PSU
FSP105-KLAM1
DE SC RIP TIO N
This series of AC /DC switching power supplies are for 90-
105 watts of continuous output power. They are enclosed
in a 94V-1 rated polyphenylene-oxide case with an
IEC 320/C 14 inlet to mate with interchangeable cord for
world-wide use. All models meet EN55011, EN55022 and
FC C class B emission limits, and are designed for medical
applications.
FEATURES
Low safety ground leakage current
Wide input range 90 to 264 VAC
OVP, OC P, OTP protection
C ompliant with C EC and Energy Star Efficiency level V
requirements
* No load power consumption less than 0.5 W
* Average active efficiency greater than 87%
Optional output connectors
W ATTAGE
Wa tta g e :
105W
DIMENSION
Dimension:
146.2mm(L) x 75.2mm(W) x
39.0mm(H)
INPUT SPECIFICATION
Input Range:
Input Frequency:
Input Current:
Leakage Current:
90-264 Vdc
47-63 Hz
1.4A(rms) for115VAC ,
0.7A(rms) for230VAC
200 µA max. @ 264 VAC ,63
Hz
SAFETY STANDARD APPAOVAL
OUTPUT SPECIFICATION
Ripple & Noise:
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500 us after a
25% step load change
Over Current
Protected to short circuit
Protection:
conditions
ENVIRONMENTAL SPECIFICATION
T E MP.Ra ng e :
Operating Temperature:0°C to
+60°C
Storage Temperature: -40°C to +
85℃
OUTPUT ELECTRICAL
Single Output:
32 V/ 3.5 A
NO TE S
Ripple and noise is maximum peak to peak voltage value measured at output within 20 MHz bandwidth, at rated line
voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the
output.
MECHANICAL SPECIFICATION
This content is subject to change, please refer to specification for m ore detail.