English
Language : 

FSP060-2K30M1 Datasheet, PDF (1/1 Pages) FSP TECHNOLOGY INC. – Medical and ITE approvals
Medical PSU
FSP060-2K30M1
DE SC RIP TIO N
This series of compact, open PC B constructed, AC -DC
switching power supplies are capable of delivering 37.5-64
watts of continuous output power at convection
cooling.They operate at 90-264 VAC input voltage without
the need of voltage ion, and are suited for medical,
information technology and industrial applications.
Approval to both EN60601-1 and EN60950-1 safety
standards improves design-in time and reduces end
equipment compliance costs.
FEATURES
Medical and ITE approvals
C ompact size 2” x 4” x 1.18”
Single, dua l and triple outputs
Wide-range input 90-264 VAC
Low earth leakage current
Level B emissions
RoHS compliant
W ATTAGE
Wa tta g e :
55W
DIMENSION
Dimension:
101.6mm(L) x 50.8mm(W) x
30.0mm(H)
INPUT SPECIFICATION
Input Range:
Input Frequency:
Input Current:
Leakage Current:
90-264 Vdc
47-63 Hz
1.3A(rms) for100VAC ,
0.7A(rms) for240VAC
150 µA max. @ 264 VAC ,63
Hz
SAFETY STANDARD APPAOVAL
OUTPUT SPECIFICATION
Ripple & Noise:
Maximum excursion of 4% or
better on all models,
recovering to 1% of final
value within 500 us after a
25% step load change
Over Current
All outputs protected to short
Protection:
circuit conditions.
GENERAL SPECIFICATION
Efficiency:
80~88% typical except PM42-31-3A
and PM42-31-5A at 75% typical
Inrush Current: 30A @ 115VAC , or 60A @ 230VAC ,
at 25"C cold start
ENVIRONMENTAL SPECIFICATION
T E MP.Ra ng e :
Operating Temperature:-10°C to
+70°C
Storage Temperature: -40°C to +
85℃
MT BF :
400,000 hours at fulll load at 25"C
ambient, calculated per MIL-HDBK-
217F
*Output Voltage and C urrent Rating
Ripple-Noise(R-P) mV
Regulation Load %
Output Max.(A)
Output Min.(A)
+5V
100mV
±3%
8A
0.5A
+12V1
120mV
±5%
3A
0.1A
NO TE S
1. Safety approvals are for PC B form only. To order unit with cover fitted, change suffix “A” to “C ”.
2. Maximum current of output #1 of multi-output models can be 8 A at 5 C FM forced air provided by user.
3. It is rated at 37.5 W maximum at convection cooling or 47.5 W maximum at 5 C FM forced air cooling by user.
4. The output voltages of a multiple output model may go outside of the stated tolerance when an output load current is
out of stated limits. All models may be operated at no-load without damage.
5. Ripple and noise is maximum peak to peak voltage value measured at output within 20 MHz bandwidth, at rated line
voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the
output.
MECHANICAL SPECIFICATION
This content is subject to change, please refer to specification for m ore detail.
FSP reserve the right to change the content without prior notice