HD600 Datasheet, PDF (7/9 Pages) Extech Instruments Corporation. – Digital Datalogging Sound Level Meter | |||
|
◁ |
ANALOG OUTPUTS
The meter is equipped with an analog output feature. The analog
output jack is located on the meterâs side panel and requires a 3.5mm
stereo phono plug. (1-AC Output, 2-DC Output, 3- Ground)
AC analog output
The AC output is a retransmitted representation of the meterâs measurement (note that
the output factors in the meterâs frequency weighting selection âAâ or âCâ).
⢠Output voltage: 1Vrms at full scale of the selected range.
⢠Output impedance: 100⦠approx.
DC analog output
The DC output signal reflects the frequency weighting selected (âAâ or âCâ).
⢠Output voltage: 10mV (±1mV) per displayed dB.
⢠Output impedance: 1K⦠approx.
Calibration
Frequent calibration is recommended and is often required by noise
standards and directives.
1. Turn the meter ON
2. Put the meter in the âAâ weighting mode
3. Put the meter in the âSLOWâ response mode
4. Place the calibrator onto the microphone.
5. Turn the calibrator ON.
6. Adjust the meterâs CAL potentiometer located on the side panel
so that the meterâs display matches the output of the calibrator
(typically 94dB or 114dB).
Battery Replacement
1. The
battery icon will appear on the display when the battery needs replacing.
2. Turn the meter OFF and slide the rear battery cover off.
3. Install the 9V battery and replace the cover.
7
HD600 V1.2 8/07
|
▷ |