CALIBRATION
The following verification and calibration procedures require the 33% and 75% RH reference
bottles which are supplied with the Model RH355 kit.
Accuracy Verification
Checking the 33% or 75% RH Calibration
1. Insert meter’s sensor into the 33% or 75% salt reference bottle
2. Check the reading after a minimum of 10 minutes
3. Verify that the reading is within the accuracy specification
Relative Humidity Calibration (33% and 75%)
33% Calibration
1. Turn the meter Off.
2. Insert the meter sensor into the 33% salt reference bottle.
3. Allow the meter one hour to stabilize.
4. While holding down the [Min/AVRST] button, press and hold the [ON/OFFSET] button
for 2 seconds to enter calibration mode.
5. “32.8%” (reading may vary depending on operating temperature) will flash on the
LCD screen. After 30 minutes, the flashing will stop to indicate that the 33%
calibration is complete. Note: To exit from the calibration procedure before
completion, pressing the [ON/OFFSET] button will turn the meter off.
6. Proceed to the 75% Calibration below.
75% Calibration
1. Insert the meter sensor into the 75% salt reference bottle.
2. Allow the meter one hour to stabilize
3. Press and hold the [Min/AVRST] button for 2 seconds to enter the 75% calibration.
4. “75.3%” (reading may vary depending on operating temperature) will flash on the
LCD. After 30 minutes, the flashing will stop to indicate that the 75% calibration is
complete.
5. Turn power off. Calibration is now complete.
Note: Frequently check the water level in the salt bottles. Both the 33% and 75%
bottles should be completely saturated with some water present. The bottles are
sealed and not refillable.
RS-232 PC Communications Interface
The meter is equipped with an RS-232 PC interface jack (3.5mm phono) for connection to
a PC. The cable to connect the meter to the PC is included in the optional 407752 data
acquisition kit. The kit includes WindowsTM compatible software that allows the user to
store readings in a text file and display real-time measurements in a series of selectable
formats. For more information or specific operating instructions, refer to the User Guide
included with the 407752 kit or contact Extech Instruments.
5
RH350/355 V2.1 8/08