English
Language : 

EX810 Datasheet, PDF (5/13 Pages) Extech Instruments Corporation. – 1000 Amp Clamp Meters with IR Thermometer
Specifications
Function
Range &
Resolution
AC Current
50/60 Hz
True RMS on
EX820 & EX830
400.0 AAC
1000 AAC
DC Current 400.0 ADC
EX830 only
1000 ADC
400.0 mVAC
AC Voltage
50/60Hz
True RMS on
EX820 & EX830
4.000 VAC
40.00 VAC
400.0 VAC
600 VAC
400.0 mVDC
4.000 VDC
DC Voltage 40.00 VDC
400.0 VDC
600 VDC
400.0Ω
EX810
Accuracy
(% of reading + digits)
EX820
± (2.8% + 8d)
± (2.5% + 8d)
EX830
± (2.5% + 8d)
± (3.0% + 8d)
± (2.8% + 5d)
± (2.8% + 5d)
-----
-----
± (1.5% + 10d)
-----
-----
± (1.0% + 10d)
± (2.5% + 5d)
± (2.8% + 5d)
± (1.0% + 10d)
± (1.8% + 8d)
± (1.5% + 5d)
± (1.5% + 5d)
± (2.5% + 8d)
± (0.8% + 2d)
± (2.0% + 5d)
± (0.8% + 2d)
± (2.0% + 5d)
± (0.8% + 2d)
± (1.5% + 2d)
± (1.5% + 2d)
± (1.5% + 2d)
± (2.0% + 2d)
± (1.0% + 4d)
± (2.0% + 2d)
± (1.0% + 4d)
± (2.0% + 2d)
± (1.0% + 4d)
4.000kΩ
40.000kΩ
Resistance
400.0kΩ
± (1.5% + 2d)
± (1.5% + 2d)
± (1.5% + 2d)
4.000MΩ
± (2.5% + 3d)
± (2.5% + 3d)
± (2.5% + 3d)
40.00MΩ
± (3.5% + 5d)
± (3.5% + 5d)
± (3.5% + 5d)
4.000nF
± (5.0% + 30d)
± (5.0% + 30d)
± (5.0% + 30d)
40.00nF
± (5.0% + 20d)
± (5.0% + 20d)
± (5.0% + 20d)
400.0nF
Capacitance 4.000µF
± (3.0% + 5d)
± (3.0% + 5d)
± (3.0% + 5d)
40.00µF
400.0µF
± (4.0% + 10d)
± (4.0% + 10d)
± (4.0% + 10d)
4.000mF
± (10% + 10d)
± (10% + 10d)
± (10% + 10d)
40.00mF
unspecified
unspecified
unspecified
4.000kHz
± (1.5% + 2d)
± (1.5% + 2d)
± (1.5% + 2d)
Frequency
Sensitivity: 100V (<50Hz); 50V (50 to 400Hz); 5V (401Hz to 4000Hz)
5