English
Language : 

407735 Datasheet, PDF (2/4 Pages) Extech Instruments Corporation. – Digital Sound Level Meter
Specifications
Display
Display update rate
Microphone
Measurement Bandwidth
Dynamic range
Measurement Range
Frequency weighting
Accuracy / Resolution
Maximum Hold decay time
Response time
Built-in calibration check
AC Analog output
DC Analog output
Power
Battery life
Operating temperature
Operating humidity
Dimensions/weight
3-1/2 digit (2000 count) LCD
0.5 seconds
0.5” Electret condensor
31.5Hz to 8KHz
65dB
35 to 130dB (Low: 35 to 90; High: 65 to 130dB)
‘A’ and ‘C’ (selectable)
± 2.0dB @1kHz / 0.1dB
<1dB / 3min
Fast: 125 milliseconds / Slow: 1 second
1KHz internal sine wave @ 94dB
0.65VAC rms (full scale); 600Ω output impedance
10mVDC / dB; 100Ω output impedance (approx).
9V Battery
50 hours (typical); Low battery indicator alerts user
32 to 104oF (0 to 40oC)
10 to 90% RH
9.45 x 2.68 x 1” (240 x 68 x 25mm) / 6.75oz (210g)
Meter Description
1. Microphone
2. LCD display
3. POWER OFF & RANGE select switch
4. RESPONSE & HOLD select switch
5. A/C weighting and Calibration switch
6. Calibration adjust screw for 94dB
7. Reset key (resets max hold reading)
8. AC/DC analog output 3.5mm phone jack
9. Battery compartment and Tripod mount are located
on back of meter
Measurement Considerations
1. Wind blowing across the microphone adds extraneous noise to the measurement.
Use the supplied windscreen to cover the microphone when applicable.
2. Calibrate the instrument before each use if possible. Especially if the meter has not
been used for a long period of time.
3. Do not store or operate the instrument in areas of high temperature or humidity.
4. Keep meter and microphone dry.
5. Avoid severe vibration when using the meter.
6. Remove the battery when the meter is to be stored for long periods of time.
2
407735 Ver. 2.1 04/2006