English
Language : 

EX613 Datasheet, PDF (12/14 Pages) Extech Instruments Corporation. – 400Amp True RMS AC/DC Clamp Meter
Specifications
Function
AC Current
50/60 Hz
True RMS
DC Current
AC/DC μA
Current
AC Voltage
50/60 Hz
True RMS
DC Voltage
Resistance
Capacitance
Frequency
(clamp)
Range
Resolution
400.0 AAC
0.1A
40.00 AAC
0.01A
400.0 ADC
0.1A
40.00 ADC
0.01A
400.00μA
0.01μA
4000.0μA
0.1μA
400.0 mVAC
0.1mV
4.000 VAC
0.001V
40.00 VAC
0.01V
400.0 VAC
0.1V
600 VAC
1V
400.00 mVDC
0.01mV
4.0000 VDC
0.0001V
40.000 VDC
0.001V
400.00 VDC
0.01V
600.0 VDC
0.1V
400.00Ω
0.01Ω
4.0000kΩ
0.0001kΩ
40.000kΩ
0.001kΩ
400.00kΩ
0.01kΩ
4.0000MΩ
0.0001MΩ
40.000MΩ
0.001MΩ
400.00nF
0.01nF
4000.0nF
0.1nF
400.00µF
0.01µF
4000.0µF
0.1µF
40.000mF
0.001mF
400.00Hz
0.01Hz
Sensitivity: 5Arms minimum
Accuracy
(% of reading + digits)
±(1.5% + 5 digits)
±(1.5% + 5 digits)
DC: ±(1.0% + 2 digits)
AC: ±(1.5% + 2 digits)
±(1.0% + 20 digits)
±(2.0% + 5 digits)
±(0.1% + 2 digits)
±(1.0% + 2 digits)
±(0.8% + 20 digits)
±(0.8% + 4 digits)
±(2.5% + 10 digits)
±(5.0% + 40 digits)
±(3.0% + 10 digits)
±(3.5% + 10 digits)
±(5.0% + 10 digits)
±(1.0% + 3 digits)
12
EX613 V2.2 07/08