STEP 8 Set calibrator to temperature Th and adjust “S” potentiometer for 20.00 mA output ± 0.01 mA.
STEP 9 Repeat Step 6 and Step 7 until both points are in tolerance.
STEP 10 Turn off power and remove wires. Mark transmitter with the new range.
TABLE 1 TL OFFSET LINKS GROUP X, Y & Z
TYPE K °C
-120 to -99
-98 to -59
-58 to -23
-22 to 9
10 to 40
41 to 72
73 to 104
105 to 130
TYPE T °C
-150 to -105
-104 to -62
-61 to -24
-23 to 9
10 to 40
41 to 71
72 to 99
100 to 120
TYPE J °C
-130 to -96
-95 to -58
-57 to -23
-22 to 9
10 to 40
41 to 72
73 to 102
103 to 130
SOLDER LINKS
-
X
Y
X, Y
Z
X, Z
Y, Z
X, Y, Z
TABLE 2 TS SPAN LINKS GROUP A, B, C & D
TYPE K °C
75 to 148
149 to 214
215 to 280
281 to 341
342 to 401
402 to 464
465 to 527
528 to 587
588 to 646
647 to 710
711 to 752
753 to 836
837 to 899
90 to 966
967 to 1035
1036 to 1075
TYPE T °C
75 to 137
138 to 189
190 to 239
240 to 284
285 to 327
328 to 371
372 to 400
-
-
-
-
-
-
-
-
-
TYPE J °C
75 to 102
103 to 170
171 to 225
226 to 283
284 to 341
342 to 409
410 to 464
465 to 521
522 to 578
579 to 642
643 to 692
693 to 743
744 to 793
794 to 851
852 to 900
901 to 1000
SOLDER LINKS
-
A
B
A, B
C
A, C
B, C
A, B, C
D
A, D
B, D
A, B, D
C, D
A, C, D
B, C, D
A, B, C, D
NOTE: Max recommended temperature
for type J input is 760 ºC.
Page 13
ALSO AVAILABLE:
Smart In Head Temperature Transmitters
DIN Rail Mounted Temperature Transmitters
Panel & Field Temperature Indicators
Temperature Probes
Trip Amplifiers
Signal Conditioners
And many other products
For further information on all products:
Status Instruments Ltd, Green Lane Business Park, Tewkesbury, Glos. GL20 8DE
Tel: +44 (0)1684 296818
Fax: +44 (0)1684 293746
Email: sales@status.co.uk