English
Language : 

PS-AN5 Datasheet, PDF (3/6 Pages) List of Unclassifed Manufacturers – Application Note PS-AN5 MediaGauge™ (Model MGA-9V) Digital Pressure Gage
SSI Technologies – Application Note PS-AN5
MediaGauge™ (Model MGA-9V) Digital Pressure Gage
Product Overview
To see the maximum peak pressure the unit has measured:
Press and hold down the LIGHT/P-H button for 3 seconds.
The maximum pressure reading will be blinking. If the
maximum pressure read was over the unit’s maximum
pressure range, o.L will be blinking on the display.
To return to the current pressure reading: Press the
LIGHT/P-H button and hold down for 3 seconds.
The RESET/CLEAR button is used to reset the gage to
whatever pressure is currently being applied to the port and
to clear the stored maximum peak pressure reading.
To reference your MediaGauge™ digital pressure gage to the
input port: Press and hold the RESET/CLEAR button for 3
seconds. The MediaGauge™ MGA-9V will now be
reference to your input port. For example, if you have 20
psi on the display and then you press and hold the
RESET/CLEAR button for 3 seconds the display will show
0.0 psi. A measurement of 30 psi would then read as 10 psi
on the display.
To reset the MediaGauge™ digital pressure gage: Make sure
the input port is either disconnected or connected to a 0 psi
input. Press and hold down the RESET/CLEAR button for 3
seconds.
To clear the maximum peak pressure reading: Press and hold
down the LIGHT/P-H button for 3 seconds until the
maximum pressure reading appears on the LCD (it will be
blinking). Release the LIGHT/P-H button. Press and hold
down RESET/CLEAR button for 3 seconds.
Installation/Mounting
1) Mount the MediaGauge™ MGA-9V digital pressure
gage on a suitable (1/4” NPT) female fitting. Use
Teflon tape or pipe dope to seal the threads.
2) Do no over tighten. Torque to 150 in lbs +/- 1 in lb.
Electrical Specifications
Supply Voltage
Response Time
Accuracy
LCD Resolution (30 PSI)
LCD Resolution (100; 300 PSI)
Proof Pressure
Burst Pressure
Operating Temperature Range
Storage Temperature Range
9V Battery
< 1 ms
±1.0 FS
0.01
0.1
3 X Full Scale
10 X Full Scale or 15,000
psi whichever is less
-10° to 60°C (14° to 140°F)
-20° to 85°C (-4°to 185°F)
SSI TECHNOLOGIES, INC.
Controls Division
2643 West Court Street
Janesville, WI 53548-5011
Phone: (608)758-1500
Fax: (608)758-2491
Copyright November 6, 2008
SSI Technologies Inc.
All Rights Reserved
Revision 3
3