English
Language : 

OVA5000 Datasheet, PDF (2/2 Pages) List of Unclassifed Manufacturers – Single measurement, all-parameter analysis of devices upto 150 m in length
OVA 5000
PARAMETER
Wavelength Range:
Wavelength:
Standard Resolution
Accuracy1
Repeatability
Optical Phase Error:
Loss Characteristics:
Dynamic range2
Ripple
Resolution
Insertion loss accuracy
Return loss accuracy
Chromatic Dispersion:
Accuracy
Group Delay:
Range3
Accuracy
Loss range2
PMD:
Range3
Accuracy – First Order
Accuracy – Second Order
Loss range2
PDL:
Extinction ratio
Accuracy
Measurement Timing:
Laser sweep rate
All-parameter measurement rate4
Fully specified measurement time5
Real-time mode update rate6
Maximum Device Length (including leads):
Transmission
Reflection
Physical
Weight (Processor not Included)
Case Size (W X D X H)
FAST MODE*
AVERAGING MODE*
1270-1340 or 1525-1610
1.6
1.6
±1.5
±1.5
±0.1
±0.1
±0.05
±0.0075
60
±0.05
±0.05
±0.1
±0.2
80
±0.01
±0.002
±0.05
±0.1
±10
±5
6
6
±0.2
±0.1
45
60
6
±0.03 (100pm steps)
±0.15 (30pm steps)
±10
40
6
±0.08
±2
50
40
±0.05
50
±0.03
70
70
30
30
12
55
1
NA
150
150
75
75
16.24
35.8
473 X 420 X 206
18.62 X 16.54 X 8.08
UNITS
nm
pm
pm
pm
radians
dB
dB
dB
dB
dB
ps/nm
ns
ps
dB
ns
ps
ps2
dB
dB
dB
nm/s
ms/nm
s
Hz
meters
meters
kg
lbs
mm
inches
1 Accuracy maintained by an internal NIST-traceable HCN gas cell.
2 80, 60 and 50 dB dynamic ranges in ‘Averaging Mode’ for IL, GD and PMD are with the
“High Dynamic Range Averaging” option installed and enabled.
3 Specifies the total device impulse-response duration that may be captured.
4 Rate calculated from combined laser sweep and analysis time per scan.
5 Measurement with full specification (see note 2) over Fast Mode: 40 nm range, and
Averaging Mode: 2.5 nm range. Excludes calibration time.
6 For 2.5 nm scan range.
Fast Mode: no averaged calibration scans, 4 averaged measurement scans, 30 pm
resolution bandwidth, 8 m device length (accuracies verified using NIST certified
artifacts except for IL). High dynamic range option enabled.
Averaging Mode: 4 averaged calibration scans, 64 averaged measurement scans,
30 pm resolution bandwidth, 8 m device length (accuracies verified using NIST
certified artifacts except for IL). High dynamic range option enabled.
*results are typical
CLASS 1 LASER PRODUCT
©Luna Innovations Incorporated, Roanoke, VA
All rights reserved.
Optical Vector AnalyzerTM is a trademark of Luna Innovations Incorporated.
LT0VA5000 REV. 005 02/13/2014