L200-TC Datasheet, PDF (2/2 Pages) List of Unclassifed Manufacturers – LABFACILITY L200 USB HIGH PRECISION DIGITAL THERMOMETER/DATA LOGGER
Specification at an ambient temperature of 20°C
Input / Ranges
Thermocouple to IEC 584
Type J
Type K
Type T
Type E
Type N
Type R
Type S
Type B
-200°C to 750°C
-200°C to1200°C
-200°C to 350°C
-200°C to 900°C
0°C to 1300°C
0°C to 1760°C
0°C to 1760°C
300°C to 1800°C
Pt100 to IEC751, 3 wire
-200°C to 850°C
Note: all inputs are non-isolated and sensors must be of
insulated construction.
Accuracy
Thermocouples J K T E & N
better than +/- 0.1°C +/-0.1% of
range -100°C to span (Zero to span
Type N) +/-0.15% of range
-101 to -200°C (J K T & E)
Thermocouples R S & B
better than+/-0.1°C +/-0.15%
of range
Linearisation
±0.05°C
Pt100 range
better than ±0.05°C ±0.1% of range
Zero drift
±0.01% of span per °C
Span drift
Display
Display resolution
±0.01% of span per °C
LCD, backlight
Thermocouple ranges 0.1°C
Pt100 range 0.01°C
User interface
Front panel key for channel No. or
auto-scan selection. PC software for
all other functions.
Indication
Reference junction
compensation for
thermocouples
Self calibration
Channel No., measured temperature
(°C or °F)
Automatic, accurate reference
junction compensation is incorporated
for thermocouple ranges
User facility incorporated.The
instrument auto-calibrates on every
A/D cycle *
Sensor open circuit
detection & indication
Ambient operating
temperature
Storage temperature
Display
Input Terminations
PC Interface
Power supply
Logging interval
Upscale indication
Independent alarms should be used
for Process safety if required
0 to 50°C
-20°C to 70°C
LCD with backlight
8 x thermocouple: mini connectors
8 x Pt100, terminal blocks
USB
6Vdc (5.5-9.0V) via universal mains
adaptor (supplied) 120-250V 50/60Hz
5 seconds to 1 hour
Specifications may be subject to change
On-board memory
PC software
512 sets of readings
Supplied as standard on CD-ROM
Remote control & measure:-
Log readings to file
Download to PC
Logging, calibration
* The integral, self-calibration facility for the thermocouple
version is a rapid and convenient method for on-site
calibration and does not require any additional equipment
other than the special, external link (optional).
Self-calibration of Pt100 ranges is quickly and conveniently
performed using plug-in precision resistors (optional).
Traceable calibration can be achieved by the user conveniently
and without recourse to a accredited Laboratory if there is
access to a certified DVM; this can be used to measure the
L200 internally generated calibrated Voltage via the “cal port”
presented externally to the instrument case. Considerable time
and cost saving are achieved by this method.
Ordering
Thermocouple input x8 L200-TC–00
Pt100 input x8
L200-PT–00
Custom calibration details on application
Standard accessories
The L200 is supplied with a power
supply adaptor, USB lead, PC software,
and instruction manual (on CD).
L200-TC includes external link.
LABFACILITY LIMITED www.labfacility.co.uk
SOUTHERN UK & EXPORT DIVISION: Units 5,6 & 7, Block K, Southern Cross Industrial Estate, Shripney Road, Bognor Regis,
West Sussex PO22 9SE Export Sales: tel: +44(0)1243 871287 fax: +44(0)1243 871281 email: exportsales@labfacility.co.uk
Southern UK Sales: tel: +44(0)1243 871280 fax: +44(0)1243 871281 email: southernsales@labfacility.co.uk
NORTHERN UK DIVISION: Eden Place, Unit 3b Outgang Lane, Dinnington, Sheffield S25 3QT
Northern UK Sales: tel: +44(0)1909 569446 fax: +44(0)1909 550632 email: northernsales@labfacility.co.uk
Data Ref: 050107DS/A Issue 4