English
Language : 

9150 Datasheet, PDF (2/2 Pages) List of Unclassifed Manufacturers – Excellent display resolution
Single Output Programmable DC Power Supplies
Models 9150, 9151, 9152 & 9153
L Rear panel
Temperature controlled
cooling fan
User accessible fuse
Remote Interface
The power supplies can be remotely controlled
from any PC through a USB (virtual com)
interface, allowing the user to program and
monitor all parameters through easy to use SCPI
commands. The power supplies come with a USB
to TTL serial converter. The average command
processing time is 35 ms.
Serial interface connector for RS232 or
USB communication
Extra Features
The 9150 series’ digital port offers a variety of
configurations. The port provides Digital Input,
external Trigger and Remote Inhibit (RI)
functionality.
Remote Sense and digital
port functionality
The RI mode can be used for turning several
power supplies On/Off simultaneously. External
triggering can be used in combination with List
mode.
Specifications
Output ratings
( 0 °C~40 °C)
Load regulation
±(% of output+offset)
Line regulation
±(% of output+offset)
Programming resolution
Readback/meter resolution
Front panel
setting resolution
Programming accuracy
12 months (25 °C ± 5 °C) ±(% of output+offset)
Readback/ meter accuracy
12 months (25 °C ± 5 °C) ±(% of output+offset)
Ripple & noise
(20 Hz ~20 MHz)
Temperature coefficient,
(0 °C~40 °C) ±(% of output+offset)
Readback temperature coefficient,
±(% of output+offset)
Transient response (for a change from 0 to
50% of maximum rated current)
DVM accuracy
DVM resolution
Milliohm meter accuracy
AC input
Weight
Dimensions (W x H x D)
Standard accessories
Optional accessories
2
v042616
9150
9151
9152
9153
0 - 5.2 V
0 - 60 A
<0.01% + 2 mV
<0.1% + 10 mA
<0.02% + 0.1 mV
<0.1% + 1 mA
0.1 mV
1 mA
0.1 mV
1 mA
0.1 mV
1 mA
<0.02%+2 mV
<0.1%+30 mA
<0.02%+1.5 mV
<0.05%+15 mA
≤4 mVp-p
15 mArms
<0.02%+2 mV
<0.1%+30 mA
<0.02%+2 mV
<0.1%+20 mA
100 µs for output to
recover to within 75 mV
0 - 20 V
0 - 27 A
0 - 30 V
0 - 18 A
0 - 60 V
0-9A
<0.01% + 1 mV
<0.1% + 5 mA
<0.01% + 1 mV
<0.1% + 2 mA
<0.02% + 1 mV
<0.01% + 1 mA
<0.02% + 1 mV
<0.01% + 0.1 mA
1 mV
1 mA
1 mV
0.1 mA
0.1 mV
0.1 mA
0.1 mV
0.1 mA
0.5 mV
1 mA
0.5 mV
1 mA
<0.02%+6 mV
<0.1%+15 mA
<0.02%+12 mV
<0.05%+10 mA
<0.02%+3 mV
<0.05%+10 mA
<0.02%+6 mV
<0.05%+5 mA
≤4 mVp-p
5 mArms
≤5 mVp-p
3 mArms
<0.02%+5 mV
<0.1%+15 mA
<0.02%+10 mV
<0.05%+5 mA
<0.02%+5 mV
≤0.05%+10 mA
<0.02%+10 mV
≤0.05%+5 mA
120 µs for output to
recover to within 75 mV
100 µs for output to
recover to within 50 mV
50 µs for output to
recover to within 50 mV
0~12V range: 0.02%+2mV
0~40V range: 0.02%+3mV
0~12V range: 0.1mV
0~40V range: 1mV
0.1% (for voltage and current ≥10% of full scale)
0.3% (for voltage and current ≥3% of full scale)
115 V/220 VAC ± 10%, 47 to 63 Hz
63.9 lbs, (29 kg)
16.88” x 3.47” x 18.06” (429 x 88 x 459 mm)
Three-Year Warranty
User manual, power cord, USB to TTL serial converter, application software installation disk, and certificate of calibration
IT-E151 rack mount kit
www.bkprecision.com