English
Language : 

SLG Datasheet, PDF (1/7 Pages) RUBYCON CORPORATION – CONDUCTIVE POLYMER ALUMINUM SOLID ELECTROLYTIC CAPACITORS
SLG Liquid Flow Meter Series
Ultra High Pressure Flow Sensor for LC Applications
 Ultra-low Flow Sensing in the Sub-µl/min Range
 Operating Pressure up to 1200 bar (17500 psi)
 40 ms Response Time
Product Summary
The SLG Liquid Flow Meter enables precise, bi-
directional measurements of dynamic liquid flow rates
in the nl/min to low µl/min range. This device has been
designed for ultra-high pressure applications in Liquid
Chromatography Systems (UHPLC). Excellent bio-
compatibility and chemical resistance is ensured by the
use of fused silica, stainless steel and PEEK for all
wetted parts.
Interface Options
Digital
- I2C-Bus
- RS485-Bus
- USB Cable
Analog
- Voltage Output (0-10 V)
- Additional operation
modes
For more information on communication, please refer
to page 2 of this document.
1 Sensing Performance
Table 1: Model specific performance of SLG (all data for medium H2O, 23°C)
Parameter
SLG-0025
SLG-0075
SLG-0150
Unit
Precision Extended
H2O Full scale flow rate
H2O Sensor output limita
1500
5000
20000
8000
1700
5500 2000/22000b 10500
nl/min
nl/min
Accuracy below full scale
(whichever error is larger)
10
10
20
5
% of m.v.c
0.5
0.5
n/a
0.5
% of full scale
Repeatability below full scale
(whichever error is larger)
0.6
0.6
1.2
0.5
% of m.v.
0.06
0.03
n/a
0.05
% of full scale
Temperature coefficient
0.3
0.3
(additional error / °C; whichever is larger)
0.03
0.03
0.3
0.15
% m.v. / °C
n/a
0.015 % full scale / °C
Mounting orientation sensitivityd
-
-
-
<0.4
% of full scale
Flow detection response time 63
40
40
120
40
ms
Response time on power-up
120
ms
Operating temperature
+10…+50
°C
Ambient storage temperaturee
-10…+60
°C
Operating pressure
1200 (17500)
500 (7250)
Bar (psi)
aFlow rate at which the sensor output saturates. See Section 2 for performance between full scale and saturation point
bLower/upper limit
cMeasured value
dNo dependence on the mounting orientation for SLG-0025 and SLG-0075. SLG-0150: maximum additional offset when mounted vertically.
eNon-condensing, flow path empty
www.sensirion.com
Version 4 – September 2015