Operating Information
The P5150 Probe is a high impedance probe with 50X
attenuation and a compensation range of 10 to 25 pF. The
probe is designed for use with Tektronix TPS2000 series
floating oscilloscopes with high input impedance (1 M Ω).
Connect the probe as shown in the illustrations below.
Compensating the Probe
Due to variations in oscilloscope input characteristics,
the low-frequency compensation of the probe may need
adjustment after you move the probe from one oscilloscope
channel to another.
1. Connect the probe to the oscilloscope channel that you
plan to use for your measurements.
2. Connect the probe to the probe compensation output
terminals on the oscilloscope front panel.
WARNING. To avoid electric shock, only connect to the
Probe Comp signal on the oscilloscope when making
this adjustment. To avoid electric shock, only use the
insulated adjustment tool when making compensation
adjustments.
3. Push AUTOSET or otherwise adjust your oscilloscope
to display a stable waveform.
4. Adjust the trimmer in the probe until you see a perfectly
flat-top square wave on the display. (See illustration.)
Standard Accessories
The probe includes the accessories shown below.
WARNING. Do not substitute accessories from other
products for use with this probe. Only use the
accessories that are included with this probe.
To avoid electric shock when using the probe or
accessories, keep your fingers behind the finger guard
of the probe body and away from the shaded area shown
in the accessory illustrations below.
Item
Description
Small hook tip (TASH)
Use this tip to access test
points in tight spaces. Screw
the hook tip onto the probe tip
and then clamp the hook onto
the circuit.
Rating: 2300 V CAT I
1000 V CAT II
Reorder Tektronix part number
013-0388-xx
Large hook tip (TALH)
Screw the hook tip onto the
probe tip and then clamp the
hook onto the circuit.
Rating: 2300 V CAT I
1000 V CAT II
Reorder Tektronix part number
013-0389-xx
Common leads
(6 in and 18 in) (TACL)
Slide the lead over the probe
head and snap it into place.
Connect the banana plug
end directly to your circuit
common, or use the crocodile
clip included with the probe.
Reorder Tektronix part
numbers:
6 inch: 196-3526-xx
18 inch: 196-3527-xx
Item
Description
Crocodile clip
Attach the clip to the banana
end of the common lead and
then to your circuit common.
Reorder Tektronix part number
344-0461-xx
Common spring (TACS)
Use this accessory to limit
aberrations on high frequency
signals caused by common
lead path inductance.
Bend the spring to reach
nearby common connections
(~1 in).
Reorder Tektronix part number
214-5299-xx
Color bands
Use these bands to identify
the oscilloscope channel at the
probe head.
Reorder Tektronix part number
016-1886-xx (5 pairs)
Adjustment tool
Use only this insulated tool for
compensation adjustments.
Reorder Tektronix part number
003-1433-xx
Optional Accessories
The accessories shown below are available for the probe
and are rated ≤30 V unless indicated otherwise.
Accessory
BNC to Tip Adapter,
Unterminated
Rubber Spring Tip
Part
number
013-0291-xx
206-0060-xx