English
Language : 

310 Datasheet, PDF (1/1 Pages) List of Unclassifed Manufacturers – Ultra Low Bias Current Varactor Bridge Operational Amplifiers
Data Sheet
Digital Milli-Ohm Meter
Model 310
The model 310 Digital Milli-Ohm Meter is used
to ensure continuity and integrity of a wire,
cable, conduit or any electrical connection. The
310 has a display resolution of 100 micro-ohms
and has a professional four wire Kelvin test lead
set included to ensure accurate readings. The
heavy duty case has a rubber seal to make the
unit water resistant and a convenient shoulder
strap.
Features & Benefits
I Four wire Kelvin lead measurements
I Over-voltage and over-temperature protection
I 5 ranges with 100 µΩ max. resolution
I Water resistant case with shoulder strap
I Auto Power Off
I IEC/EN 61010-1 / CE
I Included: 4-wire Kelvin leads, carrying case
with shoulder strap and batteries
Specifications
Electrical
Measuring ranges
Accuracy
Test current
Test current accuracy
Protection fuses
Safety
Mechanical
Bump test
Vibration test
Drop test
Impact test
Rated environmental
conditions
Weight
Dimensions (WxHD)
310
0-200.0 mΩ in steps of 100 µΩ
0-2000 mΩ in steps of 1 mΩ
0-20.00 Ω in steps of 10 mΩ
0-200.0 Ω in steps of 100 mΩ
0-2000 Ω in steps of 1 Ω
±0.5% of reading ±2 digits over the operating
temperature range -15° C to +55° C,
with the supplied test leads.
1 mA => 2000 Ω range
10 mA => 200 / 20 Ω ranges
100 mA => 2000 m / 200 mΩ ranges.
±0.3%
Supply: 1.5 A, HBC, 5 x 20 mm, DIN
Current: 1 A, HBC, 5 x 20 mm, DIN
Voltage: 0.5 A, HBC, 5 x 20 mm, DIN
LVD BS EN 61010-1
EMC BS EN 50081-1, BS EN 50082-1
IEC68-2-29
IEC1010, clause 8.3
IEC1010, clause 8.4
IEC1010, Clause 8.2
Indoor use
Pollution degree 2
Altitude up to 2000 meter
Relative humidity 80% max.
Ambient temperature 0°C~40° C
One Year Warranty
3.4 lbs (1.542 kg)
9.82” x 4.33” x 7.48” (250 x 110 x 190 mm)
Applications
The model 310 Digital milli-ohm meter, with its measuring range of 100 µΩ
to 2000 Ω, is suitable for a wide range of applications such as:
- Measuring the winding resistance of electric motors, generators and transformers
- Bond testing in mines, aircraft, railways, ships, domestic and industrial wiring
installations
- Measuring the ring main continuity testing in industrial and domestic wiring i
installations
- Measuring resistance in electronic equipment such as shunts, pcb tracks, switch
and relay resistance
- Checking compression joints on overhead lines
- Testing and maintenance of switchboard /sub-station equipment on such items
as fuses, joints, contacts and bonds
Technical data subject to change
© B&K Precision Corp. 2012
v0913122012
www.bkprecision.com