English
Language : 

2708B Datasheet, PDF (1/2 Pages) List of Unclassifed Manufacturers – Magnetic hanger
Data Sheet
Auto Ranging True RMS Tool Kit® DMM
Model 2708B
Specifications
2708B
model
TRUE RMS
Auto/ Manual Ranging
Select between Auto or
Manual ranging
The 2708B is a full-featured auto ranging true RMS DMM that combines
performance, value and functionality. The 3400 count LCD has a 34 segment bar
graph that simulates analog responsiveness. The 2708B can measure resistance
values up to 34 MΩ. With the auto power-off-feature you won’t have to worry
about accidental battery consumption. Value-packed features make this meter a
must in every "Tool Kit®".
Features
I Magnetic hanger
I True RMS
I Auto power off
I DC voltage to 1000 V
I Resistance to 34 MΩ
I Diode test
I One-handed operation
I Auto/manual ranging
I Measures up to 10 A (AC & DC)
I AC voltage to 750 V
I Continuity test
Bar Graph
The 34 segment bar graph
simulates an analog display by
updating at a rate of 12 times
per second.
DC Volts
Ranges
340 mV, 3.4 V, 34 V, 340 V, 1000V
Resolution
0.1 mV, 1 mV, 10 mV, 100 mV, 1 V
Accuracy
±(1.0 % rdg + 2 dgts)
Input Impedance
340 mV: >100 MΩ; 3.4 V: 10 MΩ; 34 V to 1000 V: 9.1 MΩ
Overload Protection
1000 VDC or 750 VACrms
AC Volts (50 H z-500 H z) True RMS
Crest Factor
<3
Ranges
3.4 V, 34 V, 340 V, 750 V
Resolution
1 mV, 10 mV, 100 mV, 1 V
Accuracy
3.4 V & 34 V ranges: ±(2.0 % rdg + 8 dgts)
340 V & 750 V ranges 50 - 500 Hz: ±(2.0 % rdg + 8 dgts)
Input Impedance
3.4 V: 10 MΩ; 34 V to 1000 V: 9.1 MΩ
Overload Protection
1000 VDC or 750 VACrms
DC Current
Ranges
340 uA, 3400 uA, 34 mA, 340 mA, 10 A
Resolution
0.1 uA, 1 uA, 10uA, 100 uA, 10 mA
Accuracy
340 uA to 340 mA ranges: ±(2.0 % rdg + 2 dgts)
10 A range: ±(3.0 % rdg + 3 dgts)
Input Protection
0.5 A/500 V & 10 A/600 V fast blow ceramic fuses
10A Input
10 A for 60 sec max. followed by a 10 minute cooling period
AC Current (50 Hz-500 Hz) True RMS
Crest Factor
<3
Ranges
340 uA, 3400 uA, 34 mA, 340 mA, 10 A
Resolution
0.1 uA, 1 uA, 10uA, 100 uA, 10 mA
Accuracy
340 uA to 340 mA ranges: ±(2.5 % rdg + 10 dgts)
10 A range: ±(3.5 % rdg + 10 dgts)
Input Protection
0.5 A/500 V & 10 A/600 V fast blow ceramic fuses
10A Input
10 A for 60 sec max. followed by a 10 minute cooling period
Resistance
Ranges
340 Ω, 3.4 kΩ, 34 kΩ, 340 kΩ, 3.4 MΩ, 34 MΩ
Resolution
0.1 Ω, 1.0 Ω, 10 Ω, 100 MΩ, 1 kΩ, 10 kΩ
Accuracy
340 Ω to 340 kΩ ranges: ±(1.5 % rdg + 4 dgts)
3.4 MΩ range: ±(2.5 % rdg + 4 dgts)
34 MΩ range: ±(5.0 % rdg + 5 dgts)
Open Circuit Voltage (typical) -0.45 VDC (-1.2 VDC on 340 Ω range)
Overload Protection
500 VDC/ACrms
Diode Test
Test Current
1.0 mA (typical)
Resolution
10 mV
Accuracy
±(3.0 % rdg + 3 dgts)
Audible Indication
<0.25 V
Open Circuit Voltage
3.0 VDC (typical)
Overload Protection
500 VDC/ACrms
Continuity
Audible Indication
Less than 35 Ω
Response Time
500 ms
Overload Protection
500 VDC/ACrms
General
Display
3 3/4 digit, 3400 count LCD with 34 segment analog bar graph
Polarity
Automatic, positive implied, negative polarity indication
Overrange
OL or -OL is displayed
Low Battery Indication
Battery symbol will display when battery voltage drops below
operating levels
Measurement Rate
2 times per second (12 times per second for analog bar graph)
Auto Power Off
Approximately 10 minutes
Operating Environment 0ºC to 50ºC at <70 % relative humidity
Storage Environment
-20ºC to 60ºC at 0 to 80 % relative humidity
Accuracies
Stated accuracy at 23ºC +5ºC at <75 % relative humidity
Temperature Coefficient 0.1 x (specified accuracy) per ºC. (ºC to 18ºC, 28ºC to 50ºC)
Power
Single standard 9 volt battery, NEDA 1604, JIS 006P, IEC 6F22
Battery Life
150 hours typical
Dimensions
165 mm (H) x 78 mm (W) x 42.5 mm (D)
Weight
Approximately 10 oz. (285 g) with battery and hanger
Three-Year Warranty
Supplied Accessories: User manual, one set of test leads, one spare fuse, and 9V battery (installed)
Technical data subject to change
© B&K Precision Corp. 2013
Page 1 of 2
www.bk precision.com
Tel.: 714.921.9095