English
Language : 

1PT100-KN0815 Datasheet, PDF (1/1 Pages) List of Unclassifed Manufacturers – CERAMIC WIRE WOUND PRTDS ARE SUITABLE
1 Pt100 KN 0815
The KN Series Ceramic Wire Wound PRTDs are suitable for
general applications requiring temperature stability.
Applications: Industrial resistance thermometers, especially
in chemical, power generation plants and analytical
equipment.
Construction: A platinum coil is sealed inside a high purity
aluminum oxide ceramic body. Lead wires are shear force
resistant and assure proper connection to extension leads
and cables.
Models
Description
Tolerance Order No.
IEC 60751
1Pt100 KN 0815
L
W0.3
W0.15
W0.1
32.206.463
32.206.464
+2
8- 0
32.206.465
Dimensions
mm
D
d
l
1.5±0.15 0.20±0.01 10.0±0.5
Self Heating
0°C (K/mW)
Response time
Water current Air stream
V=0.4m/s V=3m/s
t 0.5
t 0.9
t t 0.5
0.9
0.28
0.2 0.5 6.7 21.8
Technical Specification
Nominal resistance: 100 Ohm @ 0 °C
Temperature range: W0.3 (Class B) = -196 to +660 °C
Temperature
coefficient:
W0.15 (Class A) = -196 to +600 °C
(Heraeus exceeds IEC 60751: -100 to +450 °C)
W0.1 (Class 1/3 B) = -100 to +350 °C
Tc = 3850 ppm/K
Leads:
Palladium-gold alloy
Insulation resistance
after assembly:
> 100 MOhm @ 25 °C
Measuring current: 1 mA
Tolerance class:
- According to IEC 60751:2008
- Other standards and narrower
tolerances are available on request
Temperature stability: Excellent long-term stability
Also available:
- Platinum-gold alloy
- Different temperature coefficients
(3916 ppm/K - old JIS)
- Extension leads
- Two separated coils can be embedded
in one ceramic body
The measuring point is located at 8 mm from the end of the sensor body
Heraeus Sensor Technology USA
1901 Route 130
North Brunswick, NJ 08902
Phone 732-940-4400 Fax 732-940-4445
Email info.hst-us@heraeus.com
www.hst-us.com
Status: HSTUSA 10/09