JA31106 Datasheet, PDF (3/7 Pages) List of Unclassifed Manufacturers – One-Decimal Clinical Thermometer
JA31106
Pad Description
Pad No.
1 ~ 13
14
15
16
Pad Name
S1 ~ S13
VCC
CAP1
CAP2
17
CLFH
18
LOWC
19
VSS
20
SC
21
RF
22
RS
23
VDD
24
PSW
25
TEST
26
BZSEL
27
STSEL1
28
STSEL2
29
ML
30
31
32
33
34
35 ~ 37
FEVEL
OSCI
OSCO
BZ1
BZ2
COM1 ~
COM3
I/O
Function
O LCD segment driver
O Generated 3V voltage
O For positive voltage pumping, NMOS open drain output
O For positive voltage pumping, inverter output
Selection of °C or °F
I CLFH= Open or VDD: °C
CLFH= VSS: °F
I/O For the supply voltage detector ; output?
— Negative power supply
I/O Common point, NMOS open drain
O Connect to reference resistor, PMOS open drain
O Connect to sensor resistor, PMOS open drain
— Positive power supply
I Power ON/OFF input, toggle active
For the test of production.
I TEST=Open or VDD: LCD display the highest value
TEST=VSS: LCD display the real time value
Buzzer output frequency selection
I BZSEL=Open or VDD: 8kHz
BZSEL=VSS: 6kHz
Stable time selection (STSEL1, STSEL2 is bonding option)
ST
STSEL1
STSEL2
I
4 sec
8 sec
VSS
VDD or Open
VDD or Open
VSS
16 sec VDD or Open VDD or Open
32 sec
VSS
VSS
Memory function option
I
ML=Open or VDD: Without memory function
ML=VSS: With memory function. The last measurement value
will be displayed after power on.
Fever function option
I FEVEL=Open or VDD: With fever function
FEVEL=VSS: Without fever function
I For system oscillator input
O For system oscillator output
O Buzzer output 1
O Buzzer output 2
O LCD common driver
Absolute Maximum Ratings
Supply voltage.......................... 0V to 2.0V
Input voltage............. VSS–0.5V to VDD+0.5V
Operation Temperature......... –20°C to +75°C
Storage Temperature.......... –55°C to +125°C
*Note: Stress above those listed under “Absolute Maximum Ratings” may cause permanent damage to the
device. These are stress ratings only. Functional operation of this device at these or any other conditions
above those indicated in the operational sections of this
specification is not implied and exposure to absolute maximum rating conditions for extended periods may
affect device reliability.
TELΚ886-3-5770755
FAXΚ886-3-5770756
7-3
E-mailΚsales@jaztek.com.tw
Ver. 1.1
Web-siteΚwww.jaztek.com.tw