English
Language : 

MF-500 Datasheet, PDF (1/2 Pages) List of Unclassifed Manufacturers – SPECIFICATION SHEET FOR CO SENSOR
MEMBRAPOR
SPECIFICATION SHEET FOR CO SENSOR TYPE CO/MF-500
PERFORMANCE CHARACTERISTICS
PHYSICAL CHARACTERISTICS
Nominal Range
Maximum Overload
Expected Operation Life
Output Signal
Inboard Filter
Resolution
Temperature Range
Pressure Range
Pressure Coefficient
t90 Response Time
Relative Humidity Range
Typical Baseline Range (pure
air, 20°C)
Maximum Zero Shift (+20°C
to +40 °C)
Expected Long Term Output
Drift
Recommended Load Resistor
Bias Voltage
Repeatability
Output Linearity
0 – 500 ppm
2000 ppm
2 years in air
70 ± 20 nA/ppm
To remove H2S and
SO2
1 ppm
- 20 °C to + 50 °C
Atmospheric ± 10 %
No data
< 25 sec
15 % to 90 % R.H.
non-condensing
- 2 ppm to +3 ppm
9 ppm equivalent
< 2 % signal
loss/month
10 Ohm
Not required
< 2 % of signal
Linear
Weight
Position Sensitivity
Storage Life
Recommended Storage
Temperature
Warranty Period
~ 5.4 g
None
Six months in
container
5 °C – 20 °C
12 months from date
of dispatch
Miniature-Size Outline Dimensions
BOTTOM VIEW
Reference
90°
Sensing
Counter
18 mm
13.5 mm
PCD
SIDE VIEW
20 mm
CROSS-SENSITIVITY DATA
Interfering Gas
Concentration
Reading
H2
100 ppm
< 40 ppm
H2S
10 ppm
~ 0 ppm
5 mm
Pin Projection
1.5 mm Pins
on 13.5 mm PCD
Performance data conditions:
20 °C, 50% RH and 1013 mbar
APPLICATIONS
Safety and Environmental Control
For Portable Gas Detectors
REV.: 2/2008
Phone: +41 43 311 72 00
Fax : +41 43 311 72 01
Email: info@membrapor.ch
www.membrapor.ch
Page 1 of 2
MEMBRAPOR AG
Rautistrasse 164
CH-8048 Zürich
Switzerland