IA296
ULTRA LOW NOISE
TRUE INSTRUMENTATION
MEDICAL ISOLATION AMPLIFER
FEATURES
• UL Component Recognized
• Ultra Low Noise
• Failure of Any Internal Component Results in <10µA Bias
Current
• Excellent CMRR: 170 dB; 160 dB with 5KΩ Source
Imbalance
• External Sync Capability
The Model IA296 Isolation Amplifier features very high input /
output isolation and very low noise. It can operate at common mode
input voltages up to 5000 VDC continuous; CMRR is 170 dB with
balanced source impedance, and 160 dB with 5KΩ source imbalance.
Input voltage noise is held to 0.3µV, 10Hz – 1kHz rms. and current
noise to 4pA, 0.5Hz – 1kHz rms. Input bias current is only 200pA, and
on the failure of any component is limited to 10µA. An independent
±12VDC, 10mA supply in the input section with the same voltage
isolation as the amplifier, is used to power optional external circuitry.
BLOCK DIAGRAM IA296
+IN
-IN
IN REF
+Vs
COMMON
-Vs
+
+
Full Wave
-
Modulator
-
-
+
+
Current
Limiter Guard
-
Drive
Shield Guard
Modulator
Driver
DC-DC
Converter
Demod-
ulator
Filter
HI OUT
LO OUT
Demodulator
Driver
Oscillator
SYNC
+Vs
COMMON
SEE NOTE
6
5
2.53 (64.2) 4
MAX.
3
2
1
HI OUT 11
+IN
-IN
4 - 40 N/C LO OUT 10
IN REF 2 PLACES COM 9
+Vs
Vs COM
-Vs
SYNC 8
+15 7
BOTTOM VIEW
NOTE:
INSERT to Vs COM 300 VRMS MAX.
CHARACTERISTICS
(Typical, @ 25°C, Vs = +15 VDC unless otherwise noted.)
GAIN (Non-inverting)
ACCURACY
Vs. Temperature
Nonlinearity
Differential Input Impedance
Common Mode Input Impedance
INPUT VOLTAGE
Max. Safe Differential Input
Pulse, 10ms Duration
Max. Safe Common Mode Input (DC cont.)
INPUT BIAS CURRENT
Vs. Temperature (0 to + 70°C)
SAFETY CURRENT LIMITS
INPUT NOISE (5K Unbalance)
Voltage .05Hz to 100Hz, pp
Voltage 10Hz to 1kHz, rms.
Current .05Hz to 1kHz, rms.
COMMON MODE REJECTION
Input to Output, CMV = 115 VAC, 60Hz
Balanced Source Impedance
5KΩ Source Imbalance
Input to Reference
CMV = 10V p-p, 60Hz
5KΩ Source Imbalance
OUTPUT
Range (50KΩ Load)
Output Impedance
Max. CMV, Outputs to Power Common
Offset Voltage Referred to Input
Vs. Temperature (0° to 70°C)
Vs. Supply Voltage
FREQUENCY RESPONSE
Small Signal, - 3 dB
Full Power, 6V pp
Overload Recovery
ISOLATED POWER OUTPUTS
Voltage, ±10mA Load
Accuracy
Current
NOMINAL WEIGHT
POWER SUPPLY, SINGLRE POLARITY
Voltage, Rated Performance
Current, Quiescent
TEMPERATURE RANGE
Rated Performance
Storage
MATING SOCKET
10V / V
±2%
±. 015%/°C
±0.1%
3 X 108Ω || 2.2nF
3 X 1010Ω || 10pF
±0.5V min.
±30V rms.
±500V peak
±5000VDC
±200pA
±5pA/°C
10µA max.
1.5µV
0.3µV
4pA
170 dB
160 dB
120 dB
±5V
1 KΩ
± 30 V peak
± 5mV
± 100µV / °C
± 20µV / %
1kHz
500Hz
20mS
±12VDC
±10%
± 10mA
8 Ounces
±14.5 to +16 VDC
+ 40 mA
0°C to +70°C
-25°C to +85°C
S132
For 10kHz bandwidth specify P/N IA297 when consulting factory.
.63 (16.0)
MAX.
3.53 (89.66)
MAX.
MODEL IA296
.20 (5.08)
MIN.
.04 (1.02)
DIA.