LPS40-M_09 Datasheet, PDF (1/3 Pages) Emerson Network Power – 40 Watts Medical safety approvals | |||
|
Embedded Power for
Business-Critical Continuity
LPS40-M Series
40 Watts Medical
Total Power:
Input Voltage:
# of Outputs:
40-55 Watts
85-264 VAC
120-300 VDC
Single
Rev. 9.8.09_28
LPS40-M Series
of 3
Special Features
⢠Medical safety approvals
⢠Universal input
⢠3â x 5â footprint
⢠Less than 1U high
⢠Remote sense on main output
⢠Low output ripple
⢠Adjustable 5V output
⢠Overvoltage protection
⢠Overload protection
⢠110 KHz switching frequency
⢠LPX40 enclosure kit available
Safety
⢠VDE 0750/EN60601-1
(IEC601)
⢠UL
UL2601E182560
⢠CSA CSA 22.2 No.601.1
LR109492
⢠CE Mark (LVD)
Electrical Specifications
Input
Input raÂnge:
85 - 264 VAC; 120 - 300 VDC
Frequency:
47 - 440 Hz
Inrush current:
< 18A peak @ 115 VAC; <36A peak @ 230 VAC, cold start @ 25 °C
Input current:
1A max. (RMS) @ 115 VAC
Efficiency:
70% typical at full load
EMI filter:
Safety ground
leakage current:
FCC Class B conducted; CISPR 22 Class A conducted;
EN55022 Class A conducted; VDE 0878 PT3 Class A conducted
< 75µA @ 50/60 Hz, 264 VAC input
Output
Maximum power:
40 W for convection; 55 W with 30CFM forced air
Adjustment range:
-5, +10% minimum
Hold-up time:
20ms @ 40 W load, 115 VAC nominal line
Overload protection: Short circuit protection on all outputs
Case overload protected @ 110 - 145% above peak rating
Overvoltage protection: 5V output; 5.7 to 6.7 VDC. Other outputs 10% to 25% above nominal
output
Remote sense:
Compensates for 0.5 V lead drop minimum.
Will operate without remote sense connected.