English
Language : 

MDS-300ADB48AA Datasheet, PDF (6/13 Pages) Delta Electronics, Inc. – Medical AC-DC Enclosed Power Supply
Technical Datasheet
Medical AC-DC Enclosed Power Supply
48V 300W with 5V/0.5A Standby / MDS-300ADB48 AA
Protections (for both 48 Volt and 5Volt Standby outputs, unless otherwise indicated)
Overvoltage (max)
Over load / Over current (max)
Main output 125% of rated voltage, Latch Mode
Standby Output 125% of rated voltage, Latch Mode
Main output 160% of rated current
Standby Output 1.06A max
Hiccup Mode(Non-Latching, Auto-Recovery)
Over Temperature
Short Circuit
Latch Mode for main output
Hiccup Mode,
(Non-Latching, Auto-Recovery)
Reliability Data
MTBF(Minimum) at 115Vac, 210W, 35 °C
Convection Air Flow
Operating life(Minimum) at 115Vac, 210W, 25°C Convection Air Flow
800 Khrs based on Telecordia SR-332
26,280 hrs
Safety Standards / Directives
Medical Safety
IEC60601-1 2nd , 3rd and 3.1rd edition CB report
IEC60601-1 edition 3.1rd (2012), EN60601-1 (2006) +
A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14,
ANSI/AAMI ES60601-1:2005/(R)2012
ITE Safety
CE
Environmental
IEC60950-1 CB report
TUV60950-1
UL60950-1+CAN/CSA60950-1
GB4943.1-2011, GB9254-2008, GB17625.1-2003
MDD Directive 93/42/EEC
RoHS Directive 2011/65/EU Compliant
Galvanic Isolation
Input to/Output (2XMOPP) 4000 Vac
Input to/Ground(1XMOPP) 1500 Vac1)
Output to/Ground(1XMOPP) 1500 Vac ( Type BF application rated)
1) PSU can support PoE applications with Primary to FG 2500Vac test.
6
All parameters are specified at 25°C ambient unless otherwise noted.