MDS-400APB24AA Datasheet, PDF (2/12 Pages) Delta Electronics, Inc. – Medical AC-DC Open Frame Power Supply
Medical AC-DC Open Frame Power Supply
24V/400W 3”x5” High Density MDS-400APB24 AA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
Nominal Input Frequency
Input Frequency Range
Input Current (max)
Input Surge Voltage (max)
Full load Efficiency (typ.)
Standby Power (max)
Inrush Current (max)
Input-PE(protective earth) leakage current(max)
Output-PE(protective earth) leakage current for Type BF application (max)
Power Factor (min)
1) NC: normal condition, SFC: single fault condition
100-240Vac
90-264Vac
50-60Hz
47-63Hz
5.5A
300Vac for 100ms
92% @ 115Vac/60Hz
93% @ 230Vac/50Hz, Reference Fig.1
0.5W (only standby working with Inhibit signal high)
@ 115Vac/60Hz, 230Vac/50Hz
40A@230Vac, cold start
0.1mA @ NC, 0.3mA @ SFC1)
0.1mA @ NC, 0.5mA @ SFC 1)
0.95 @ 115V/50Hz, 230V/50Hz, full load
Leakage Current
Input-PE Leakage Current
100Vac/60Hz(Typ)
Normal Condition 18.3uA
Single Fault Condition 33.7uA
Output-PE Leakage Current forType BF application
Normal Condition 29.4uA
Single Fault Condition 43.5uA
264Vac/60Hz(Typ) Delta Limit
44.6uA
100uA max
91.9uA
300uA max
87.3uA
130.5uA
100uA max
500uA max
IEC60601-1 Limit
5000uA max
10000uA max
100uA max
500uA max
100.000
90.000
80.000
70.000
60.000
115Vac
230Vac
50.000
25%
50%
75%
Output Load
Fig.1 Efficiency versus output load
100%
2
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 08)