English
Language : 

MDS-400ADB12AA Datasheet, PDF (2/11 Pages) Delta Electronics, Inc. – Medical AC-DC Enclosed Power Supply
Medical AC-DC Enclosed Power Supply
12V 400W 3.5”x5.5” High Density MDS-400ADB12 AA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
Nominal Input Frequency
Input Frequency Range
Input Current(max)
Input Surge Voltage (max)
Full load Efficiency (typ.)
Standby Power (max)
Inrush Current (max)
Input-PE(protective earth) leakage current(max)
Output-PE(protective earth) leakage current for Type BF application (max)
Power Factor (min)
1) NC: normal condition, SFC: single fault condition
100-240Vac
90-264Vac
50-60Hz
47-63Hz
5.5A
300Vac for 100ms
89.5% @ 115Vac/60Hz
91.5% @ 230Vac/50Hz, Reference Fig.1
0.5W (only standby working with Inhibit signal high)
@ 115Vac/60Hz, 230Vac/50Hz
40A@230Vac, cold start
0.1mA @ NC, 0.3mA @ SFC1)
0.1mA @ NC, 0.5mA @ SFC 1)
0.95 @ 115V/50Hz, 230V/50Hz, full load
Leakage Current
Input-PE Leakage Current
100Vac/60Hz(Typ)
Normal Condition 17.5uA
Single Fault Condition 32.9uA
Output-PE Leakage Current for Type BF application
Normal Condition 28.5uA
Single Fault Condition 42.9uA
264Vac/60Hz(Typ) Delta Limit
43.5uA
100uA max
90.7uA
300uA max
86.7uA
128.6uA
100uA max
500uA max
IEC60601-1 Limit
5000uA max
10000uA max
100uA max
500uA max
100.00
90.00
80.00
70.00
60.00
115Vac
230Vac
50.00
25%
50%
Output Load (400W)
75%
Fig.1 Efficiency versus output load
2
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 03)
100%