English
Language : 

MDS-200APB24-AA Datasheet, PDF (2/11 Pages) Delta Electronics, Inc. – Medical AC-DC Enclosed Power Supply
Medical AC-DC Enclosed Power Supply
24V 200W 2”x4” High Density / MDS-200APB24 AA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
Nominal Input Frequency
Input Frequency Range
Input Current (max)
Input Surge Voltage (max)
Full load Efficiency (typ.)
Standby Power (max)
Inrush Current (max)
Input-PE(protective earth) leakage current(max)
Output-PE(protective earth) leakage current for Type BF application (max)
Power Factor (min)
1) NC: normal condition, SFC: single fault condition
100-240Vac
90-264Vac
50-60Hz
47-63Hz
2.5A
300Vac for 100ms
92% @ 115Vac/60Hz
93% @ 230Vac/50Hz, Reference Fig.1
0.5W @ 115Vac/60Hz, 230Vac/50Hz
60A@230Vac, cold start
0.1mA @ NC, 0.3mA @ SFC1)
0.1mA @ NC, 0.5mA @ SFC 1)
0.95 @ 115V/50Hz, 230V/50Hz, full load
Leakage Current
Input-PE Leakage Current
100Vac/60Hz(Typ)
Normal Condition 19.5uA
Single Fault Condition 44.8uA
Output-PE Leakage Current for Type BF application
Normal Condition 38.3uA
Single Fault Condition 43.4uA
264Vac/60Hz(Typ) Delta Limit
45.0uA
100uA max
153.0uA
300uA max
81.5uA
129.4uA
100uA max
500uA max
IEC60601-1 Limit
5000uA max
10000uA max
100uA max
500uA max
100
90
115Vac
80
230Vac
70
60
50
25%
50%
75%
Output Load
100%
Fig.1 Efficiency versus output load
2
All parameters are specified at 25°C ambient unless otherwise noted.
(November 2016, Rev. 07)