T``E`CHNICAL DATASHEET
MDS Medical AC-DC Adapter
19Volt, 150Watt / MDS-150AAS19 F/B
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
Nominal Input Frequency
Input Frequency Range
Input Current (max.)
Efficiency (typ.)
Average Efficiency (min)
Standby Power (max.)
Inrush Current (typ.)
Touch Current (max.)
1) NC: normal condition
2) SFC: single fault condition
100-240Vac
90-264Vac
50-60Hz
47-63Hz
4A @ 115Vac, 2A @ 230Vac
90.5%, Reference Fig.1
88% @ 115Vac and 230Vac
0.21W @ 115Vac and 230Vac
50A @ 115Vac, 100A @ 230Vac
0.1mA @ 264Vac NC1), 0.3mA @ 264Vac SFC2)
Output Ratings / Characteristics
Fig.1 Efficiency versus output load
Nominal Output Voltage
Total Regulation
Output Current
Output Power
Line Regulation (max)
Load Regulation (max)
Ripple & Noise (typ.)
Start-up Time (max)
Hold-up Time (min)
Dynamic Response (Overshoot & Undershoot O/P Voltage)
19Vdc
± 5%
7.89A
150W
±0.5%
±4.5%
94.4mV pk-pk @ Full load, Reference Fig. 2,
3000ms
20ms @ 115Vac&230Vac
± 5% @ 50-100% load
2
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 03)