T``E`CHNICAL DATASHEET
MDS Medical AC-DC Open Frame
18V100W / MDS-100APS18 BA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
Nominal Input Frequency
Input Frequency Range
Input Current (max)
Efficiency (typ.)
Inrush Current (typ.)
Standby Power (max)
Earth Leakage Current (max)
1) NC: normal condition
2) SFC: single fault condition
100-240Vac
90-264Vac
50-60Hz
47-63Hz
1.5A @ 100Vac, 0.75A @ 240Vac
91.1%, Reference Fig.1
30A @ 115Vac, 60 A @ 230Vac
0.3W
0.1mA @ 264Vac NC1), 0.3mA @ 264Vac SFC2)
Fig.1 Efficiency versus output load
Output Ratings / Characteristics
Nominal Output Voltage
Output Voltage Tolerance
Output Current
Output Power
Line Regulation (max)
Load Regulation (max)
Ripple & Noise (typ.)
Start-up Time(max)
Hold-up Time(min)
Dynamic Response (Overshoot & Undershoot O/P Voltage)
*Periodic and Random Deviation
18Vdc
± 1%
5.55A for convection
100W for convection
±0.5%
±1%
30mV pk-pk @ Full load, Reference Fig. 2
3000ms @ 115Vac
20ms @ 115Vac
± 3% @ 50-100% load
2
All parameters are specified at 25°C ambient unless otherwise noted.