MDS-065APS18BA Datasheet, PDF (2/10 Pages) Delta Electronics, Inc. – MDS Medical AC-DC Open Frame
TECHNICAL DATASHEET
MDS Medical AC-DC Open Frame
18V 65W / MDS-065APS18 BA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
Nominal Input Frequency
Input Frequency Range
Input Current (max)
Efficiency (typ.)
Standby Power (max)
Inrush Current (typ.)
Earth Leakage Current (max)
1) NC: normal condition
2) SFC: single fault condition
100
95
90
85
80
75
70
65
60
55
50
25%
100-240Vac
90-264Vac
50-60Hz
47-63Hz
1.5A @ 100Vac, 0.75A @ 240Vac
87.7%, Reference Fig.1
0.3W
30A @ 115Vac, 60 A @ 230Vac
0.1mA @ 264Vac NC1), 0.3mA @ 264Vac SFC2)
115Vac Efficiency
230Vac Efficiency
50%
75%
Ou tp ut loa d
Fig.1 Efficiency versus output load
100%
Output Ratings / Characteristics
Nominal Output Voltage
18Vdc
Output Current
3.61A
Output Power
65W
Output Peak Load
4.98A*
Line Regulation (max)
±0.5%
Load Regulation (max)
±1%
Ripple & Noise (typ.)
34mV pk-pk @ Full load, Reference Fig. 2
Start-up Time(max)
3000ms @ 115Vac
Hold-up Time(min)
15ms @ 115Vac
Dynamic Response (Overshoot & Undershoot O/P Voltage)
± 3% @ 50-100% load
*Output peak can last 8 sec with 10% duty cycle, and the output voltage should meet ±5% voltage regulation at normal input voltage
2
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (July 2013, Rev. 00)