English
Language : 

MDS-040APS18BA Datasheet, PDF (2/11 Pages) Delta Electronics, Inc. – MDS Medical AC-DC Open Frame
TECHNICAL DATASHEET
MDS Medical AC-DC Open Frame
18V 40W / MDS-040APS18 BA
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
Input Voltage Range
Nominal Input Frequency
Input Frequency Range
Input Current (max.)
Efficiency (typ.)
Standby Power (max.)
Inrush Current (typ.)
Earth Leakage Current (max.)
1) NC: normal condition
2) SFC: single fault condition
100-240Vac
90-264Vac
50-60Hz
47-63Hz
0.85A @ 115V
86.9%, Reference Fig.1
0.3W
30A @ 115Vac, 60A @ 230Vac
0.1mA @ 264Vac NC1), 0.3mA @ 264Vac SFC2)
Output Ratings / Characteristics
Fig.1 Efficiency versus output load
Nominal Output Voltage
Output Voltage Tolerance
Output Current
Output Power
Line Regulation (max)
Load Regulation (max)
Ripple & Noise (typ.)
Start-up Time(max)
Hold-up Time(min)
Dynamic Response (Overshoot & Undershoot O/P Voltage)
*Periodic and Random Deviation
18Vdc
± 5%
2.67A
40W
±0.5%
±1%
14.8mV pk-pk @ Full load, Reference Fig. 2
3000ms @ 115Vac
15ms @ 115Vac
± 3% @ 50-100% load
2
All parameters are specified at 25°C ambient unless otherwise noted.