English
Language : 

MDS-400ADB24AA Datasheet, PDF (1/11 Pages) Delta Electronics, Inc. – Medical AC-DC Enclosed Power Supply
Medical AC-DC Enclosed Power Supply
24V 400W 3.5”x5.5” High Density MDS-400ADB24 AA
MDS-400ADB24 AA
Highlights & Features
–Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
–Compliant with IEC 60601-1-2 4th Ed. Requirements
–Up to 400W forced air
– IT & Medical Safety Approvals
Safety Standards
General Description
CB Certified for worldwide use
Model Number:
MDS-400ADB24 AA
Unit Weight:
605g(1.34lb)
Dimensions (W x L x H):
88.9x140.0x44.4mm (3.5x5.5x1.75 inch)
Delta’s MDS-400ADB24 enclosed design offers a high energy density (11.9 Watts/in3) 3.5x5.5x1.75 inch design for Type BF patient
access medical products. With operating parameters of 90 to 264 Vac universal input voltage, temperatures of -10 to +70 degrees
centigrade, and altitudes of up to 5000 meters (16,402 feet), the design is well suited for a variety of both medical and non-medical
applications. Other features include input surge of 300 Vac, low leakage current, no-load input power < 0.5 Watt, and 500K hour MTBF.
This product is certified for EMC standards EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment, and EN
55022 for Industrial Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are
used to ensure long product life.
An enhanced feature set, includes a 5 Volt/2 amp standby output, remote on/off, remote sense, and a power good signal that are
included as part of the standard product.
The design, which has protection against shock compliant with 2XMOPP and Type BF requirements, has both medical (with risk report
available), and ITE safety approvals, including cURus(US&Canada) /TUV/GB(China); plus, CB certificates with all national deviations.
Full compliance with RoHS Directive 2011/65/EU for environmental protection is included.
Model Information (All with 5V/2A standby available):
Model Number
Input Voltage Output Voltage Forced Air Output Current
MDS-400ADB24 AA
90-264Vac 24Vdc
0-16.67A (with 16 CFM forced air)
Fan Output Voltage
Fan Output Current
12Vdc
0.05-0.6A1)
1) Fan output will be presenting when 24V main output is available
MDS
Delta Medical
power Supply
400
Max wattage in the
product Series.
400 → 400W
A
Family Code
A~ Z
D
Product Type
D: Enclosed
B
Input Type Code
B: 3pin Class I
24
Output Voltage
24 for 24V
AA
Revision code
1
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (November 2016, Rev. 03)